Software as a Medical Device (SaMD): Does Your Product Count, and What Follows

Last reviewed: . Sources opened and read on the dates shown.

Short answer: Software as a medical device (SaMD) is software intended for a medical purpose, such as diagnosing or treating disease, that works without being part of a hardware medical device. Whether your product counts depends on its intended use, which your marketing claims help set, and on statutory exclusions. Decide per function, in writing, before you build further.

This guide is for a founder or product lead at a digital health company deciding whether the product is a medical device, before a full-time CTO or regulatory hire. It draws on the statute, the regulations and FDA's own pages, opened on November 28, 2026. It is general information, not regulatory or legal advice. FDA guidance changes, so check dates before relying on any summary, this one included.

What is software as a medical device (SaMD)?

FDA uses the definition agreed by the International Medical Device Regulators Forum (IMDRF), a group of national regulators: “software intended to be used for one or more medical purposes that perform these purposes without being part of a hardware medical device” (FDA, Software as a Medical Device (SaMD), checked November 28, 2026). FDA's examples include software that processes images to help detect breast cancer (FDA, What are examples of Software as a Medical Device?, checked November 28, 2026).

Is your product a medical device? Five questions, in order

Answer for each function separately. Stop at the first clear answer and write down why.

QuestionWhat the law or FDA saysWhat your answer means
1. What is the intended use?The objective intent of the people responsible for the product. The regulation says: “This objective intent may, for example, be shown by labeling claims, advertising matter, or oral or written statements by such persons or their representatives.” (eCFR, 21 CFR 801.4, meaning of intended uses, checked November 28, 2026)Write one sentence per function: who uses it, on which patients, for what condition, and what they do with the output. Then collect every claim you make anywhere.
2. Does that use meet the device definition?A device includes an article “intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, in man or other animals” (U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 321(h), device definition (FD&C Act section 201(h)), checked November 28, 2026), or intended to affect the structure or any function of the body.No: not a device. Re-check when claims or features change. Yes: go on.
3. Does a statutory software exclusion apply?Before it reaches decision support, section 520(o) removes four kinds of function: administrative support of a health care facility; software “for maintaining or encouraging a healthy lifestyle and is unrelated to the diagnosis, cure, mitigation, prevention, or treatment of a disease or condition” (U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 360j(o), software exclusions (FD&C Act section 520(o)), checked November 28, 2026); electronic patient records that meet listed conditions; and transferring, storing or displaying lab or device data without interpreting or analyzing it.Yes, squarely, for every claim you make: not a device. Partly: go on.
4. Is it clinical decision support that meets all four criteria?This is the fifth exclusion. All four criteria in the next table must be met.Yes: not a device. Fails one: go on.
5. Is it a device function FDA says it does not intend to regulate?For many functions that meet the definition but pose minimal risk, FDA says “the FDA will exercise enforcement discretion and will not expect manufacturers to submit premarket review applications or to register and list their software with the FDA” (FDA, Device Software Functions Including Mobile Medical Applications, checked November 28, 2026). One example: helping patients self-manage a condition without specific treatment suggestions. A separate policy covers low-risk general wellness products (FDA guidance, General Wellness: Policy for Low Risk Devices, checked November 28, 2026).Yes: enforcement discretion, which FDA describes as its intent at this time (FDA, Digital Health Policy Navigator, checked November 28, 2026). The function may still be a device. No: FDA oversees it. See the next section.

The clinical decision support exclusion, in the statute's words

Section 520(o)(1)(E) excludes a function only if it passes one gate and serves three purposes. FDA's January 29, 2026 guidance calls these Criteria 1 to 4 (FDA, Clinical Decision Support Software, guidance document dated January 29, 2026 (PDF), checked November 28, 2026).

CriterionThe statute's wordsA plain check, with FDA's reading
1Excluded “unless the function is intended to acquire, process, or analyze a medical image or a signal from an in vitro diagnostic device or a pattern or signal from a signal acquisition system” (U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 360j(o), software exclusions (FD&C Act section 520(o)), checked November 28, 2026)If the function takes in a medical image, a lab-instrument signal, or a pattern such as an ECG waveform or continuous glucose monitor readings, the exclusion is out.
2Intended for “displaying, analyzing, or printing medical information about a patient or other medical information (such as peer-reviewed clinical studies and clinical practice guidelines)” (U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 360j(o), software exclusions (FD&C Act section 520(o)), checked November 28, 2026)Inputs are what clinicians tell each other: a lab result, a diagnosis, a guideline.
3Intended for “supporting or providing recommendations to a health care professional about prevention, diagnosis, or treatment of a disease or condition” (U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 360j(o), software exclusions (FD&C Act section 520(o)), checked November 28, 2026)The user is a health care professional, and the output informs their judgment without directing it. FDA: “Software functions that support or provide recommendations to patients or caregivers – not HCPs – meet the definition of a device.” (FDA, Clinical Decision Support Software, guidance document dated January 29, 2026 (PDF), checked November 28, 2026)
4Intended for “enabling such health care professional to independently review the basis for such recommendations that such software presents so that it is not the intent that such health care professional rely primarily on any of such recommendations to make a clinical diagnosis or treatment decision regarding an individual patient” (U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 360j(o), software exclusions (FD&C Act section 520(o)), checked November 28, 2026)The clinician can see what the recommendation rests on and has time to check it. FDA: “FDA does not consider software functions intended for a critical, time-sensitive task or decision to meet Criterion 4” (FDA, Clinical Decision Support Software, guidance document dated January 29, 2026 (PDF), checked November 28, 2026)

Where a function offers only one clinically appropriate recommendation but meets every other criterion, FDA says it intends to exercise enforcement discretion (FDA, Clinical Decision Support Software, guidance document dated January 29, 2026 (PDF), checked November 28, 2026).

Two caveats. Intended use comes from what you say as well as what you build: website copy, app store text, sales decks, what staff tell customers. FDA's general wellness guidance: “FDA expects that a product intended for general wellness use will have labeling (including instructions for use, user-facing claims, promotional materials, and marketing communications) that is consistent with, and does not exceed, the product’s stated intended use.” (FDA, General Wellness: Policy for Low Risk Devices, guidance document dated January 6, 2026 (PDF), checked November 28, 2026)

And your analysis is a position, not a ruling. FDA's screening tool states: “The results of this Navigator are not a formal FDA device determination for your product.” (FDA, Digital Health Policy Navigator, checked November 28, 2026) Guidance itself “does not establish any rights for any person and is not binding on FDA or the public” (FDA, Clinical Decision Support Software, guidance document dated January 29, 2026 (PDF), checked November 28, 2026). FDA, not the company, makes the formal determination.

What follows if it is a device: what to plan and budget for

If a function is a device and not under enforcement discretion, plan for six things early.

AreaWhat FDA or the rule saysWhat it means for budget and hiring
ClassificationClass I is subject to general controls, class II adds special controls, and class III requires premarket approval (eCFR, 21 CFR 860.3, definitions of class I, class II and class III, checked November 28, 2026). FDA: “Most Class I devices are exempt from Premarket Notification 510(k); most Class II devices require Premarket Notification 510(k); and most Class III devices require Premarket Approval.” (FDA, Overview of Device Regulation, checked November 28, 2026) IMDRF's four SaMD risk categories, I (lowest impact) to IV, are a different scheme (FDA, Global Approach to Software as a Medical Device, checked November 28, 2026).Search FDA's classification database for products like yours. Note the class and product code (FDA, Classify Your Medical Device, checked November 28, 2026).
Premarket routeA 510(k) shows substantial equivalence to a legally marketed predicate device. A De Novo request is for novel class I or II devices with no predicate (eCFR, 21 CFR 860.200, De Novo classification: purpose and applicability, checked November 28, 2026). Premarket approval (PMA), for class III, is the most stringent. Some class I and II devices are exempt (FDA, How to Study and Market Your Device, checked November 28, 2026).FDA's fiscal year 2027 fees: 510(k) $28,653; De Novo $191,020; PMA $636,732. Qualified small businesses pay $7,163, $47,755 and $159,183. Annual establishment registration is $13,785 (FDA, Medical Device User Fee Amendments (MDUFA): Fees, FY 2027, checked November 28, 2026). Consulting, testing and clinical studies are extra; get written quotes.
Quality system21 CFR Part 820 opens: “Current good manufacturing practice (CGMP) requirements are set forth in this quality management system regulation (QMSR).” (eCFR, 21 CFR 820.1, scope of the Quality Management System Regulation, checked November 28, 2026) FDA says the revised rule took effect on February 2, 2026 and incorporates ISO 13485:2016 (FDA, Quality Management System Regulation: Frequently Asked Questions, checked November 28, 2026). A manufacturer must: “Document a quality management system that complies with the applicable requirements of ISO 13485” (eCFR, 21 CFR 820.10, requirements for a quality management system, checked November 28, 2026). Design and development requirements cover class II and III devices and class I devices automated with computer software.Requirements, risk management, testing and change control are written and traceable. Someone owns quality, as an employee or a contractor.
Software documentationFDA's premarket software guidance sets a “Documentation Level, which is either Basic or Enhanced” (FDA, Content of Premarket Submissions for Device Software Functions, guidance document dated June 14, 2023 (PDF), checked November 28, 2026). Enhanced applies where a software failure could present a probable risk of death or serious injury. Both call for a risk management file, requirements, architecture diagrams and testing evidence.Keep these records as you build. Reconstructing them later costs engineering time.
CybersecurityFDA: “Congress granted the FDA explicit cybersecurity regulatory authority in 2022 by adding Section 524B to the FD&C Act.” (FDA, Cybersecurity in Medical Devices Frequently Asked Questions (FAQs), checked November 28, 2026) A cyber device includes software, can connect to the internet and could be vulnerable to threats. Its submission must include a plan for vulnerabilities found after launch, processes for updates and patches, and a software bill of materials (SBOM).A component inventory, a disclosure and patching process, and a named security owner.
After launchManufacturers register and list the device with FDA, and report incidents in which it may have caused or contributed to a death or serious injury (FDA, Overview of Device Regulation, checked November 28, 2026).A complaint and reporting process, and a person who runs it.

For usability testing, see our guide to human factors for SaMD and wearables.

What changes for AI that updates

FDA says: “The FDA does not regulate AI as such; it regulates medical devices, including AI-enabled medical devices.” (FDA, Artificial Intelligence-Enabled Medical Devices, checked November 28, 2026) The same questions and routes apply. What differs is change.

On terms: FDA says it uses Super Intelligence (SI) in place of AI in public documents newly issued from November 16, 2026, and that: “Existing guidance documents, regulations, Federal Register notices, and other previously issued FDA documents remain in effect as written.” (FDA’s Use of “Super Intelligence” and “SI” Under Executive Order 14434, checked November 28, 2026)

FDA software as a medical device guidance: titles, status and dates

Status and dates are as FDA's own pages showed them on November 28, 2026. Guidance is FDA's current thinking, not law. Three of these seven were reissued in January and February 2026.

TopicFDA's title, linkedStatus and date on FDA's page
Clinical decision support (FDA guidance, Clinical Decision Support Software, checked November 28, 2026)Final, January 2026. Dated January 29, 2026; supersedes a version issued January 6, 2026 (FDA, Clinical Decision Support Software, guidance document dated January 29, 2026 (PDF), checked November 28, 2026).
General wellness (FDA guidance, General Wellness: Policy for Low Risk Devices, checked November 28, 2026)Final, January 2026. Supersedes the September 27, 2019 version (FDA, General Wellness: Policy for Low Risk Devices, guidance document dated January 6, 2026 (PDF), checked November 28, 2026).
Device software functions and mobile apps (FDA guidance, Policy for Device Software Functions and Mobile Medical Applications, checked November 28, 2026)Final, September 2022.
Software content of a premarket submission (FDA guidance, Content of Premarket Submissions for Device Software Functions, checked November 28, 2026)Final, June 2023.
Cybersecurity (FDA guidance, Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions, checked November 28, 2026)Final, February 2026. Supersedes the June 27, 2025 version.
Change control plans for AI (FDA guidance, Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions, checked November 28, 2026)Final, August 2025.
AI lifecycle and submissions (FDA guidance, Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendations, checked November 28, 2026)Draft, January 2025. Not for implementation.

Check each guidance's own page, not a list. FDA's index of digital health guidances still showed the cybersecurity guidance under its 2025 title and date when we checked (FDA, Guidances with Digital Health Content, checked November 28, 2026).

Common ways teams get this wrong

These are hypothetical patterns, not accounts of real companies.

  • The disclaimer and the homepage disagree. The terms of use say the app is not a medical device. The homepage says it detects a named disease. Both are evidence of intended use (eCFR, 21 CFR 801.4, meaning of intended uses, checked November 28, 2026).
  • A clinician in the loop is treated as the whole test. A team relies on the decision support exclusion because a doctor sees the output. But the function analyzes a heart-rhythm waveform, or sends the same advice to patients. Either fails the criteria (FDA, Clinical Decision Support Software, guidance document dated January 29, 2026 (PDF), checked November 28, 2026).
  • Wellness wording, disease claim. A stress app is marketed as helping treat an anxiety disorder. FDA's guidance lists that claim as outside general wellness (FDA, General Wellness: Policy for Low Risk Devices, guidance document dated January 6, 2026 (PDF), checked November 28, 2026).
  • Enforcement discretion is reported as not regulated. A team tells investors FDA does not apply. The accurate statement is narrower: FDA does not intend to enforce at this time, and a new feature or claim can end that.
  • A web service is assumed to be different. FDA's FAQ says “a developer that is providing users access to a medical device software through a website subscription, software as a service, or other similar means, is also considered a device manufacturer” (FDA, Clinical Decision Support Software Frequently Asked Questions (FAQs), checked November 28, 2026).
  • The memo is older than the guidance. An analysis from 2025 cites a document FDA replaced in January 2026. Date every regulatory memo. Re-check it before a raise, a launch or a change in claims.

What to do first

  1. List the functions. Write a one-sentence intended use for each.
  2. Gather every claim. Website, app store listing, decks, onboarding screens, support scripts. Mark any that name a disease, a diagnosis or a treatment.
  3. Run each function through the five questions. FDA's Digital Health Policy Navigator covers the same ground in seven steps (FDA, Digital Health Policy Navigator, checked November 28, 2026).
  4. Write a short position memo. The answer for each function, the reasoning, the documents and dates relied on, and what would change it. Have a regulatory consultant or counsel review it.
  5. If the answer is unclear, ask FDA. Its Digital Health Center of Excellence takes policy questions by email (FDA, Clinical Decision Support Software Frequently Asked Questions (FAQs), checked November 28, 2026). For a formal answer, FDA says: “If you would like a formal device determination or classification from the FDA, consider submitting a 513(g) Request.” (FDA, Classify Your Medical Device, checked November 28, 2026) The fiscal year 2027 fee is $8,596, or $4,298 for a qualified small business (FDA, Medical Device User Fee Amendments (MDUFA): Fees, FY 2027, checked November 28, 2026). Device-specific feedback before a submission goes through the Q-Submission program (FDA guidance, Requests for Feedback and Meetings for Medical Device Submissions: The Q-Submission Program, checked November 28, 2026).
  6. If a function is a device, choose before the next build cycle. Change the claim or feature so it is not one, or start the quality system now.

Questions to ask a regulatory consultant or counsel

  • Which of our functions are device functions, and under which provision or guidance?
  • Which claims in our current marketing change your answer?
  • If you rely on the decision support exclusion, how do we meet each criterion, and what must the product show the clinician?
  • If you rely on enforcement discretion, which guidance and which example? What would take us out of it?
  • What class and product code do you expect, and which route? What is the predicate?
  • What evidence will our claims need: bench testing, analytical validation, a clinical study?
  • Would you ask FDA first, by 513(g) request or Pre-Submission?
  • What must the quality system cover before we write more code, and who should own it?
  • For model updates, what belongs in a predetermined change control plan?
  • Which FDA documents, by title and date, does your advice rest on?
  • What is your fee basis, and what is outside it?

Where a fractional CTO fits

A regulatory consultant or counsel owns the legal position and any submission. A technical lead owns what makes that position true in the product: which functions exist, what data they take in, what the clinician sees and how changes are controlled.

At Palo Alto Health Labs, regulatory posture, covering HIPAA and Software as a Medical Device, is part of the fractional CTO work for digital health companies. It is not legal advice and does not replace a regulatory consultant. Nobody can promise how FDA will view a product. For scoped work on this question, see SaMD planning and consulting.

Disclosure: Palo Alto Health Labs sells fractional CTO work, Fabrica and scoped build or migration work, so weigh its advice on hiring help accordingly.

See also custom healthcare software development, the EHR integration guide and the fractional CTO hub.

Frequently asked questions

What does SaMD stand for?

Software as a Medical Device. FDA uses the international regulators' definition: “software intended to be used for one or more medical purposes that perform these purposes without being part of a hardware medical device” (FDA, Software as a Medical Device (SaMD), checked November 28, 2026).

Does FDA regulate software as a medical device?

Yes, where a software function meets the device definition and no statutory exclusion applies. FDA says it applies oversight to functions whose failure could pose a risk to a patient's safety, and does not intend to enforce requirements for some lower-risk functions (FDA guidance, Policy for Device Software Functions and Mobile Medical Applications, checked November 28, 2026).

Is clinical decision support software a medical device?

Not if it meets all four criteria in section 520(o)(1)(E) (U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 360j(o), software exclusions (FD&C Act section 520(o)), checked November 28, 2026). Under FDA's January 2026 guidance, a function that analyzes images or signals, advises patients, or supports time-critical decisions does not meet them (FDA, Clinical Decision Support Software, guidance document dated January 29, 2026 (PDF), checked November 28, 2026).

Is a wellness app a medical device?

The statute excludes healthy-lifestyle software that is unrelated to diagnosing or treating a disease (U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 360j(o), software exclusions (FD&C Act section 520(o)), checked November 28, 2026). FDA also has a policy for low-risk general wellness products. A claim to diagnose or treat a disease falls outside both (FDA, General Wellness: Policy for Low Risk Devices, guidance document dated January 6, 2026 (PDF), checked November 28, 2026).

Sources

Each outside claim on this page links to the page it came from, opened and read on the date shown. Figures and terms change; check the current source before relying on one. This page is general information, not legal advice.

  1. FDA, Software as a Medical Device (SaMD). Checked November 28, 2026.
  2. FDA, What are examples of Software as a Medical Device?. Checked November 28, 2026.
  3. FDA, Digital Health Policy Navigator. Checked November 28, 2026.
  4. U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 360j(o), software exclusions (FD&C Act section 520(o)). Checked November 28, 2026.
  5. eCFR, 21 CFR 801.4, meaning of intended uses. Checked November 28, 2026.
  6. U.S. Code (Office of the Law Revision Counsel), 21 U.S.C. 321(h), device definition (FD&C Act section 201(h)). Checked November 28, 2026.
  7. FDA, Device Software Functions Including Mobile Medical Applications. Checked November 28, 2026.
  8. FDA guidance, General Wellness: Policy for Low Risk Devices. Checked November 28, 2026.
  9. FDA, Clinical Decision Support Software, guidance document dated January 29, 2026 (PDF). Checked November 28, 2026.
  10. FDA, General Wellness: Policy for Low Risk Devices, guidance document dated January 6, 2026 (PDF). Checked November 28, 2026.
  11. eCFR, 21 CFR 860.3, definitions of class I, class II and class III. Checked November 28, 2026.
  12. FDA, Overview of Device Regulation. Checked November 28, 2026.
  13. FDA, Global Approach to Software as a Medical Device. Checked November 28, 2026.
  14. FDA, Classify Your Medical Device. Checked November 28, 2026.
  15. eCFR, 21 CFR 860.200, De Novo classification: purpose and applicability. Checked November 28, 2026.
  16. FDA, How to Study and Market Your Device. Checked November 28, 2026.
  17. FDA, Medical Device User Fee Amendments (MDUFA): Fees, FY 2027. Checked November 28, 2026.
  18. eCFR, 21 CFR 820.1, scope of the Quality Management System Regulation. Checked November 28, 2026.
  19. FDA, Quality Management System Regulation: Frequently Asked Questions. Checked November 28, 2026.
  20. eCFR, 21 CFR 820.10, requirements for a quality management system. Checked November 28, 2026.
  21. FDA, Content of Premarket Submissions for Device Software Functions, guidance document dated June 14, 2023 (PDF). Checked November 28, 2026.
  22. FDA, Cybersecurity in Medical Devices Frequently Asked Questions (FAQs). Checked November 28, 2026.
  23. FDA, Artificial Intelligence-Enabled Medical Devices. Checked November 28, 2026.
  24. eCFR, 21 CFR 807.81, when a premarket notification submission is required. Checked November 28, 2026.
  25. FDA guidance, Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions. Checked November 28, 2026.
  26. FDA guidance, Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendations. Checked November 28, 2026.
  27. FDA’s Use of “Super Intelligence” and “SI” Under Executive Order 14434. Checked November 28, 2026.
  28. FDA guidance, Clinical Decision Support Software. Checked November 28, 2026.
  29. FDA guidance, Policy for Device Software Functions and Mobile Medical Applications. Checked November 28, 2026.
  30. FDA guidance, Content of Premarket Submissions for Device Software Functions. Checked November 28, 2026.
  31. FDA guidance, Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions. Checked November 28, 2026.
  32. FDA, Guidances with Digital Health Content. Checked November 28, 2026.
  33. FDA, Clinical Decision Support Software Frequently Asked Questions (FAQs). Checked November 28, 2026.
  34. FDA guidance, Requests for Feedback and Meetings for Medical Device Submissions: The Q-Submission Program. Checked November 28, 2026.