Palo Alto Health Labs All services
Guide · Regulatory

FDA Medical Device Cybersecurity and Section 524B: A Guide for Connected Wearables and SaMD

If your device includes software and can connect to the internet, FDA's cybersecurity rules almost certainly apply to your 510(k), De Novo or PMA. Section 524B of the FD&C Act, added by FDORA in December 2022, makes certain cybersecurity content a legal requirement for "cyber devices" [1][3]. FDA's premarket cybersecurity guidance explains what to submit [1][2].

Format
Guide
Read time
7 min
Checked
October 27, 2026

Facts checked October 27, 2026.

What is the current FDA guidance?

The current guidance is "Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions" (February 2026), docket FDA-2021-D-1158. It supersedes the June 27, 2025 final guidance, which had in turn updated the September 2023 version and finalized the March 2024 section 524B draft [1][4]. The title now says "Quality Management System" instead of "Quality System," in line with FDA's Quality Management System Regulation (QMSR). The QMSR amends 21 CFR Part 820 to incorporate ISO 13485:2016 by reference. It took effect February 2, 2026; the final rule was published February 2, 2024 (FR Doc. 2024-01709) [6][7].

What is a "cyber device"?

Under section 524B(c), a cyber device is a device that:

  1. includes software validated, installed, or authorized by the sponsor as a device or in a device;
  2. has the ability to connect to the internet; and
  3. contains any such technological characteristics that could be vulnerable to cybersecurity threats [2][3].

Most connected wearables, companion-app systems and cloud SaMD will meet this definition. FDA's guidance discusses edge cases [2].

What does section 524B require?

Sponsors of a 510(k), PMA, PDP, De Novo or HDE for a cyber device must:

  1. submit a plan to monitor, identify and address postmarket vulnerabilities and exploits, including coordinated vulnerability disclosure;
  2. design, develop and maintain processes giving reasonable assurance that the device and related systems are cybersecure, and make postmarket updates and patches available; and
  3. provide a software bill of materials (SBOM) covering commercial, open-source and off-the-shelf components [3].

This applies to Special and Abbreviated 510(k)s and to PMA/HDE supplements too [3].

When did it take effect?

What does FDA expect in a premarket submission?

Per the guidance [1][2]; confirm section numbers against the February 2026 PDF:

Checklist for wearable and SaMD startups

How PAHL helps

Palo Alto Health Labs helps teams design system architecture and security for connected health products:

We are not a notified body or a penetration-testing lab, and we don't guarantee clearance. Book a 30-minute scoping call.

Sources (accessed October 27, 2026)

  1. FDA, Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions (February 2026): https://www.fda.gov/regulatory-information/search-fda-guidance-documents/cybersecurity-medical-devices-quality-management-system-considerations-and-content-premarket
  2. FDA, guidance PDF: https://www.fda.gov/media/119933/download
  3. FDA, Cybersecurity in Medical Devices Frequently Asked Questions (section 524B): https://www.fda.gov/medical-devices/digital-health-center-excellence/cybersecurity-medical-devices-frequently-asked-questions-faqs
  4. Federal Register, notice of availability of the June 27, 2025 guidance (background on section 524B and FDORA section 3305): https://www.federalregister.gov/documents/2025/06/27/2025-11669/cybersecurity-in-medical-devices-quality-system-considerations-and-content-of-premarket-submissions
  5. Statute: FD&C Act section 524B, 21 U.S.C. 360n-2, "Ensuring cybersecurity of devices" (U.S. Code, 2024 edition, GovInfo): https://www.govinfo.gov/content/pkg/USCODE-2024-title21/html/USCODE-2024-title21-chap9-subchapV-partA-sec360n-2.htm
  6. Federal Register, "Medical Devices; Quality System Regulation Amendments," final rule (published February 2, 2024; effective February 2, 2026): https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments
  7. FDA, Quality Management System Regulation (QMSR): https://www.fda.gov/medical-devices/postmarket-requirements-devices/quality-management-system-regulation-qmsr

Frequently asked questions

Does 524B apply to my wellness wearable?

Only to devices in premarket submissions. A product that isn't a device isn't subject to it, but the design practices are still worth following.

Is an SBOM legally required?

Yes, for cyber devices under section 524B(b)(3).

Do device modifications need cybersecurity documentation?

Yes. Premarket submissions for changes to cyber devices are covered, and the guidance has specific recommendations.

Is the guidance binding?

Guidance is nonbinding. The section 524B requirements are statutory.

Related

Questions

Does 524B apply to my wellness wearable?

Only to devices in premarket submissions. A product that isn't a device isn't subject to it, but the design practices are still worth following.

Is an SBOM legally required?

Yes, for cyber devices under section 524B(b)(3).

Do device modifications need cybersecurity documentation?

Yes. Premarket submissions for changes to cyber devices are covered, and the guidance has specific recommendations.

Is the guidance binding?

Guidance is nonbinding. The section 524B requirements are statutory.