How FDA classifies seizure-monitoring wearables
Non-EEG wearables that flag seizures fall under 21 CFR 882.1580, "Non-electroencephalogram (EEG) physiological signal based seizure monitoring system": "A noninvasive prescription device that collects physiological signals other than EEG to identify physiological signals that may be associated with a seizure." It is Class II with special controls, product code POS.
- Clinical testing must show the device functions "as an assessment aid for monitoring for seizure-related activity."
- "Performance measurements must include positive percent agreement and false alarm rate."
- Labeling is required for both clinicians and patients or caregivers, including instructions "regarding how to set the device threshold."
EEG devices are a separate regulation, 21 CFR 882.1400, with product codes including GWQ (full-montage standard EEG), OMC (reduced-montage EEG) and OMB (automatic event detection software for full-montage EEG).
Sources: 21 CFR 882.1580 (eCFR); openFDA classification, 882.1580; openFDA classification, 882.1400. This page is a summary, not medical or regulatory advice.
Devices on record under product code POS
The table is a partial list of POS premarket records, oldest first. It may miss devices under other codes, and we have not summarized intended use or performance: read each summary on FDA's database before stating what a device detects. Clearance or a De Novo grant is not FDA approval.
| Device | Number | Applicant | Decision date |
|---|---|---|---|
| Companion | DEN140033 (De Novo) | LGCH Inc. | 2017-02-16 |
| Embrace | K172935 | Empatica | 2018-01-26 |
| Embrace | K181861 | Empatica | 2018-12-20 |
| SPEAC System | K182180 | Brain Sentinel | 2019-05-11 |
| SPEAC System | K200276 | Brain Sentinel | 2021-02-06 |
| EpiMonitor | K232915 | Empatica | 2024-02-15 |
| EpiWatch Monitoring System | K243515 | Epiwatch Inc. | 2025-03-07 |
| EpiMonitor | K250515 | Empatica | 2025-06-19 |
| NightWatch+ US | K243199 | Livassured BV | 2025-07-12 |
| Nelli (Version 7.11) | K251506 | Neuro Event Labs | 2025-11-21 |
Sources: openFDA 510(k) records, POS; checked on the FDA 510(k) database and FDA De Novo database. openFDA describes its data as unvalidated.
Clinical guidance
The ILAE and IFCN published a joint clinical practice guideline on automated seizure detection using wearable devices in 2021. Its recommendations focus on tonic-clonic seizures; read the guideline itself for the exact scope before relying on it.
What it means for startups
Our reading (editorial, not from a source):
- A seizure-flagging wearable is a device, not a wellness product: prescription, Class II, with a 510(k) path under POS.
- Design the study around the required metrics. Positive percent agreement and false alarm rate need confirmed seizure events and enough recording time to estimate false alarms.
- Labeling seizures takes expert annotation: onset, offset and seizure type, with agreement checks between reviewers.
- Claims beyond the guideline's focus need their own evidence.
Related resources
Frequently asked questions
Are seizure-detection wearables FDA-regulated?
Yes. Non-EEG seizure monitoring wearables are Class II prescription devices under 21 CFR 882.1580, product code POS.
What performance does FDA require for a seizure wearable?
The special controls require clinical testing, and performance measurements must include positive percent agreement and false alarm rate.
Which seizure wearables are on FDA's records?
Product code POS includes records for Companion, Embrace, SPEAC System, EpiMonitor, EpiWatch, NightWatch+ US and Nelli, among others. Read each 510(k) or De Novo summary for its intended use.
Scope it in 30 minutes
Book a scoping call and we'll name the smallest engagement that moves you forward. Most teams start with the fixed-scope Initial Audit.
Book a 30-min scoping callPrefer to write? Send a short note.