Key message: In published validation studies, smartwatch and smartphone screening tools picked up about 9 in 10 people with moderate-to-severe sleep apnea (pooled smartwatch sensitivity 89%), but how often a positive result was correct varied a lot between studies. A flag from a consumer device is a good reason to book a proper sleep test, not a diagnosis, and about half of the evidence comes from studies involving the device or app maker.
What the studies found
Smartwatches. Four smartwatch studies compared the watch with an overnight lab sleep study (polysomnography) at the usual threshold for moderate-to-severe sleep apnea (15 or more breathing events per hour, AHI ≥15), used a pre-set cut-off, and reported enough detail to count true and false results. Pooled together, the watches detected 89.0% of people with moderate-to-severe sleep apnea (95% CI 81.7% to 93.6%) and correctly cleared 79.9% of people without it (95% CI 51.4% to 93.7%). Detection was fairly consistent across studies. Specificity was not: the prediction interval, which estimates the range a new study might show, runs from 8.0% to 99.5%. A fifth study reported 92.3% sensitivity and 92.6% specificity, but at a cut-off chosen within its own data and against an automatically scored sleep study, so it is shown separately; adding it gives 89.2% and 82.2%. In practice, a watch that misses few cases may still raise many false alarms in some settings.
What that means for 1,000 people. As an illustration only: if 100 of 1,000 people tested had moderate-to-severe sleep apnea, a tool performing at the pooled averages would flag about 89 of them and miss about 11. It would also flag about 180 of the 900 people without it, so roughly 1 in 3 flags would be a true case. In groups where sleep apnea is less common, the share of correct flags drops further.
Phone apps. Eleven smartphone studies were found, mostly listening to breathing and snoring sounds, and some using sonar or motion. Three of them gave full counts that could be pooled. One (Tiron 2020) prints a 2x2 table. The other two were rebuilt from group sizes and rates printed in the full text, and every reported figure was reproduced. Pooled, these apps detected 87.4% of people with moderate-to-severe sleep apnea (95% CI 71.7% to 95.0%), with specificity of 82.7% (95% CI 69.7% to 90.8%). With only three studies, the prediction intervals are wide (sensitivity 26.2% to 99.3%; specificity 35.5% to 97.6%). Other phone studies not in the pool reported results in a similar range, for example 79% sensitivity with 80% specificity against a home sleep test, and 72.0% with 85.5% in a home study.
Smart rings and under-mattress sensors. These were too few, or too varied, to pool, so they are described rather than combined. In four ring studies, detection at AHI ≥15 ranged from 66% to 87%, and specificity from 79% to 100%. Two of these studies set their cut-off within their own data, which tends to flatter results. One regulatory submission tested a ring and a watch on the same 469 to 475 people: the ring detected fewer cases (77.9% vs 90.8%) but had fewer false alarms (95.9% vs 90.3% specificity). For under-mattress sensors, one lab study reported 88.0% sensitivity and 88.6% specificity at AHI ≥15. A 2026 study found that averaging a month of nights changed the balance: 77% sensitivity and 85% specificity, compared with 86% and 79% for a single night.
Regulatory decisions. Since 2024, the US Food and Drug Administration (FDA) has authorised three over-the-counter software features that assess sleep apnea risk on consumer watches and rings (product code QZW). Their public decision documents include clinical results. One feature was designed to give a notification only after repeated signs over 30 nights. It was tested against a home sleep test and had lower sensitivity (66.3%) but very high specificity (98.5%), with no notifications in participants whose AHI was under 5. This shows a deliberate trade-off: fewer false alarms in exchange for missing more cases.
How this compares with published reviews. Earlier meta-analyses report similar overall numbers. One reported pooled sensitivity of 0.87 and specificity of 0.86 for wearables at AHI >15, and another reported 86.3% and 87.3% for audio-based AI. The pooled estimates here are in line with those, with a wider spread in specificity.
Figures

Figure 1. Sensitivity (share of people with moderate-to-severe sleep apnea who were flagged) and specificity (share without it who were correctly cleared) at AHI ≥15, grouped by device type. Filled navy squares are studies included in a pooled estimate. Hollow orange circles are studies shown but not pooled. Black diamonds are pooled estimates (REML with floored Hartung-Knapp 95% CI), labelled with the pooled value. Bars are 95% confidence intervals, either exact from counts or as reported. The study marked ‡ used a cut-off chosen within its own data and is not in the primary pool.

Figure 2. Number of validation studies and people tested per device type, split by whether the device or app maker ran, funded or co-authored the study. Hatched bars are studies with no reported company role.
Pooled results at a glance (AHI ≥15)
| Group (studies) | Measure | Pooled | 95% CI (floored Hartung-Knapp) | 95% CI (Wald) | Tau (tau²), logit | q | 95% prediction interval | I² |
|---|---|---|---|---|---|---|---|---|
| Smartwatch (4) | Sensitivity | 89.0% | 81.7–93.6% | 84.9–92.1% | 0.24 (0.057) | 0.82 | 68.7–96.8% | 40% |
| Smartwatch (4) | Specificity | 79.9% | 51.4–93.7% | 63.8–90.0% | 0.78 (0.615) | 1.00 | 8.0–99.5% | 89% |
| Phone app (3) | Sensitivity | 87.4% | 71.7–95.0% | 81.4–91.6% | 0.00 (0.000) | 0.39 | 26.2–99.3% | 0% |
| Phone app (3) | Specificity | 82.7% | 69.7–90.8% | 77.4–86.9% | 0.00 (0.000) | 0.21 | 35.5–97.6% | 0% |
Caveats
- Who ran the studies. 17 of the 33 study records involve the company that makes the device or app, through funding, employment or co-authorship, and the regulatory results come from company submissions. This is normal for product validation, but independent replication is still limited. Funding and conflict-of-interest information for each study is taken as stated in the study or FDA document.
- One manufacturer dominates the watch evidence. Two of the four pooled watch studies test the same commercial feature, in different versions and settings, and a third uses a different algorithm on the same manufacturer's hardware. The two watch studies with counts that are shown separately (the earlier FDA version and Kim 2025) also test that feature, so five of the six watch estimates with 2x2 counts at AHI ≥15 use that manufacturer's devices. The pooled watch estimate therefore mostly reflects that manufacturer's devices, not all smartwatches.
- Failed or unusable recordings. Several studies lost a substantial share of data. One watch study excluded 37 of 90 participants (41%), mostly for missing or insufficient watch data; another rejected 26.5% of oximetry data; one FDA watch submission had insufficient device data on 16.7% of nights, and another reported a result for 87.9% of sessions; one phone study had no complete result for 8 of 150 people. Accuracy figures apply only to recordings that worked.
- Possible overlap. Public documents do not say whether the people in the second FDA watch submission and in Alavi 2026 overlap; the primary pool treats them as independent.
- Study populations. Most studies recruited people already referred for a sleep test, where sleep apnea is common. Accuracy is often lower in the general population, and positive results are less likely to be correct there.
- Different methods. Reference tests (lab versus home sleep tests, scoring rules), device outputs (one night versus many nights) and cut-offs differ between studies. Some cut-offs were chosen within the same data used to test them, which tends to overstate accuracy.
- Night-to-night variation. Sleep apnea severity varies between nights, so a single-night comparison can count some true results as errors, and the other way round.
- Screening, not diagnosis. None of these tools replaces a sleep test. A "no flag" result does not rule out sleep apnea, especially in people with symptoms.
- About this review. This is a rapid, non-systematic review. Searching, screening, extraction and analysis were done with the Palo Alto Health Labs Co-Scientist, an AI research assistant, with no independent double extraction by a second reviewer. The same team designed the analysis and checked its own work, so the quality checks are self-graded. Treat this as an informal summary, not a clinical guideline.
Methods
Design. Rapid non-systematic review of diagnostic accuracy studies of consumer devices for obstructive sleep apnea in adults, with FDA decision documents and published systematic reviews used for context.
Sources and dates. PubMed (NCBI E-utilities), OpenAlex (works API) and the openFDA device database, plus FDA decision PDFs from accessdata.fda.gov. All searches were run on 23 November 2026.
Search terms.
- PubMed primary studies (2015–2026; 242 records):
("sleep apnea"[tiab] OR "sleep apnoea"[tiab] OR "sleep-disordered breathing"[tiab] OR "obstructive sleep apnea"[mh]) AND (smartwatch*[tiab] OR "smart watch"[tiab] OR wristband[tiab] OR "smart ring"[tiab] OR ring[tiab] OR smartphone*[tiab] OR "mobile phone"[tiab] OR consumer[tiab] OR wearable*[tiab] OR "under-mattress"[tiab] OR "undermattress"[tiab] OR "bed sensor"[tiab] OR contactless[tiab] OR sonar[tiab] OR acoustic[tiab] OR "snoring sound*"[tiab] OR "breathing sound*"[tiab] OR "Apple Watch"[tiab] OR Fitbit[tiab] OR Oura[tiab] OR Withings[tiab] OR "Galaxy Watch"[tiab] OR "mobile app*"[tiab] OR "smartphone app*"[tiab]) AND (polysomnograph*[tiab] OR "home sleep apnea test*"[tiab] OR "home sleep test*"[tiab] OR polygraph*[tiab] OR "portable monitor*"[tiab] OR "type 3"[tiab] OR "type III"[tiab]) AND (sensitivity[tiab] OR specificity[tiab] OR "diagnostic accuracy"[tiab] OR validation[tiab] OR "area under"[tiab] OR AUC[tiab] OR AUROC[tiab]) AND 2015:2026[dp] - PubMed reviews (2020–2026; 38 records): the same first two blocks AND
(systematic[sb] OR "meta-analysis"[pt] OR "systematic review"[tiab] OR "meta-analysis"[tiab]) AND 2020:2026[dp] - PubMed coverage checks using device and brand terms:
Snorefox;"Apple Watch" AND (apnea OR apnoea);(Oura OR "smart ring") AND (apnea OR apnoea);"Galaxy Watch" AND (apnea OR apnoea);(Wellue OR Circul OR "O2Ring") AND (apnea OR apnoea);Withings AND (apnea OR apnoea);(Huawei OR Amazfit OR Xiaomi OR Garmin OR Fitbit OR Google) AND (apnea OR apnoea) AND (polysomnograph* OR "sleep test");(smartphone OR "mobile phone") AND (apnea OR apnoea) AND (sonar OR contactless OR "breathing sound*") AND polysomnograph*. A phrase search for"sleep apnea feature" OR "sleep apnea notification"returned mostly irrelevant records and was not screened. - OpenAlex (title and abstract, 2015–2026; 454 records):
("sleep apnea" OR "sleep apnoea") AND (smartwatch OR "smart ring" OR smartphone OR wearable OR "under-mattress" OR contactless OR "consumer") AND (polysomnography OR "home sleep apnea test") AND (sensitivity OR specificity OR accuracy) - openFDA: product code QZW ("over-the-counter device to assess risk of sleep apnea", 21 CFR 868.2378), plus applicant searches for major consumer device makers.
Screening. 738 records and 4 FDA documents were screened. 92 were assessed in detail, and 29 publications plus 3 FDA decisions were included (33 study records, 32 independent groups of people), along with 10 systematic reviews.
Inclusion rules. Studies were included if they met all of these:
- adults
- a consumer device or an app on consumer hardware (smartwatch, wristband, smart or consumer ring oximeter, smartphone, consumer under-mattress sensor)
- a reference of lab or home polysomnography, or a home sleep apnea test
- per-person accuracy at an AHI threshold
- published 2015–2026
Studies were excluded for any of these:
- children
- prescription medical home sleep test devices
- research-grade or modified sensors
- questionnaire-only apps
- sleep-staging or oxygen-only outcomes
- conference abstracts
- duplicate cohorts
Data extraction. Values came from open full text (PubMed Central) where available, and otherwise from abstracts or FDA documents. Where a paper did not print the 2x2 table, counts were derived from printed group sizes and reported rates. These counts were used only when they reproduced every reported accuracy figure exactly. No values were imputed. Of the three pooled phone studies, Tiron 2020 prints its test-set 2x2 table (which reproduces its reported sensitivity, specificity, PPV, NPV and accuracy). The Sommer 2024 and Le 2023 tables were rebuilt from the group sizes printed in each full text (people or nights above and below AHI 15) plus the reported rates. Each rebuilt table reproduces every accuracy figure the study reports at AHI ≥15: sensitivity, specificity, PPV, NPV and exact confidence intervals for Sommer 2024, and sensitivity and specificity, the only per-person figures reported, for Le 2023. For Alavi 2026 only the abstract was available; the analysed n of 90 and the 2x2 counts were inferred from the published Wilson confidence intervals, which those counts reproduce uniquely. Kim 2025 was moved out of the primary watch pool because its reported accuracy uses a cut-off optimised in the same data (estimated AHI 11.1) against an AI-scored sleep study; it is kept as a sensitivity analysis. The inputs used are listed in the count_source column of sources_v2.csv. Narayan 2019 was not pooled because counts implied by its reported rates (32 with and 27 without moderate-to-severe sleep apnea) conflict with its own baseline table (33 and 26). DOIs were taken from each article's own PubMed record and checked against the title and first author in Crossref.
Statistical analysis. Pooling was done only where at least 3 comparable, independent studies at the same threshold (AHI ≥15) had full counts. Sensitivity and specificity were pooled separately on the logit scale (univariate random effects). Between-study variance was estimated by restricted maximum likelihood (REML). The main 95% CI uses the floored Hartung-Knapp method: the REML standard error is multiplied by the square root of max(1, q), where q is the weighted residual variance of the study estimates, with a t distribution on k−1 degrees of freedom. Under this floor the interval is never narrower than the Wald CI. q is reported for each estimate, and Wald CIs are also shown. Other outputs:
- tau and tau² (between-study SD and variance on the logit scale)
- a 95% prediction interval (t on k−2 degrees of freedom)
- I² from Cochran's Q
A 0.5 continuity correction was planned for zero cells but was not needed. Sensitivity analyses (floored Hartung-Knapp CIs) covered three changes to the primary watch pool:
- adding Kim 2025 (in-sample cut-off, AI-scored reference; k=5): sensitivity 89.2% (95% CI 83.4–93.2%), specificity 82.2% (61.2–93.1%)
- swapping in the earlier version of the main watch feature (k=4): 85.8% (78.8–90.8%) and 78.9% (53.5–92.4%)
- dropping a study counted per wrist rather than per person (k=3): 91.0% (81.2–96.0%) and 76.8% (24.0–97.2%)
As a check, a bivariate random-effects model (logit sensitivity and specificity, unstructured between-study covariance, REML; R metafor rma.mv, Wald CIs, since Hartung-Knapp has no standard bivariate form) was fitted. For the primary watch pool it gave sensitivity 89.5% (95% CI 85.3–92.6%) and specificity 79.9% (63.6–90.0%), moving the univariate estimates by 0.5 and less than 0.1 percentage points. For the phone pool it moved them by less than 0.1 points. The main analysis used a Python equivalent of R metafor rma(method="REML", test="knha") with q floored at 1, written in Python 3 with NumPy, SciPy (REML optimisation) and statsmodels (Paule-Mandel tau² cross-check; exact Clopper-Pearson CIs for single studies), and Matplotlib for figures.
Studies pooled.
- Smartwatch: FDA 510(k) K261011 (2026), Alavi 2026, Zhou 2024, Jung 2022.
- Phone: Tiron 2020, Sommer 2024, Le 2023.
The full extraction table is in sources_v2.csv.
References
- US Food and Drug Administration (2024). De Novo Classification Request for Sleep Apnea Feature (DEN230041); applicant Samsung Electronics. https://www.accessdata.fda.gov/cdrh_docs/reviews/DEN230041.pdf
- US Food and Drug Administration (2026). 510(k) K261011 Sleep Apnea Feature (v2.0); applicant Samsung Electronics. https://www.accessdata.fda.gov/cdrh_docs/pdf26/K261011.pdf
- US Food and Drug Administration (2024). 510(k) K240929 Sleep Apnea Notification Feature (SANF); applicant Apple Inc.. https://www.accessdata.fda.gov/cdrh_docs/pdf24/K240929.pdf
- Alavi et al. (2026). Smartwatch-based detection of moderate-to-severe and high-risk obstructive sleep apnea. https://doi.org/10.1007/s44470-026-00159-8
- Kim et al. (2025). AI‑Enhanced Smartwatch AHI Estimation and AI‑Scored Polysomnography for Obstructive Sleep Apnea: Real‑World Validation. https://doi.org/10.2147/NSS.S540460
- Zhou et al. (2024). Comparison of OPPO Watch Sleep Analyzer and Polysomnography for Obstructive Sleep Apnea Screening. https://doi.org/10.2147/NSS.S438065
- Cinar Bilge et al. (2024). Scanning of obstructive sleep apnea syndrome using smartwatch: A comparison of smartwatch and polysomnography. https://doi.org/10.1016/j.jocn.2023.12.009
- Kim et al. (2022). Diagnostic Performance of Photoplethysmography-Based Smartwatch for Obstructive Sleep Apnea. https://doi.org/10.18787/jr.2022.00424
- Chen et al. (2021). A Single-Center Validation of the Accuracy of a Photoplethysmography-Based Smartwatch for Screening Obstructive Sleep Apnea. https://doi.org/10.2147/NSS.S323286
- Browne et al. (2024). Performance of a commercial smart watch compared to polysomnography reference for overnight continuous oximetry measurement and sleep apnea evaluation. https://doi.org/10.5664/jcsm.11178
- Jung et al. (2022). Performance evaluation of a wrist-worn reflectance pulse oximeter during sleep. https://doi.org/10.1016/j.sleh.2022.04.003
- Hayano et al. (2025). Detection of sleep apnea using only inertial measurement unit signals from apple watch: a pilot-study with machine learning approach. https://doi.org/10.1007/s11325-025-03255-w
- Kang et al. (2025). Usefulness of Continuous Sleep Tracking by Smartwatch in Predicting Sleep Apnea. https://doi.org/10.3342/kjorl-hns.2024.00178
- Benedetti et al. (2022). Obstructive Sleep Apnoea Syndrome Screening Through Wrist-Worn Smartbands: A Machine-Learning Approach. https://doi.org/10.2147/NSS.S352335
- Tisyakorn et al. (2024). Screening moderate to severe obstructive sleep apnea with wearable device. https://doi.org/10.1007/s11325-024-03232-9
- Zhao et al. (2022). Comparison of Ring Pulse Oximetry Using Reflective Photoplethysmography and PSG in the Detection of OSA in Chinese Adults: A Pilot Study. https://doi.org/10.2147/NSS.S367400
- Meira E Cruz et al. (2025). A wearable ring oximeter for detection of sleep disordered breathing. https://doi.org/10.1016/j.rmed.2025.108092
- Tiron et al. (2020). Screening for obstructive sleep apnea with novel hybrid acoustic smartphone app technology. https://doi.org/10.21037/jtd-20-804
- Lyon et al. (2019). Detection of Sleep Apnea Using Sonar Smartphone Technology. https://doi.org/10.1109/EMBC.2019.8857836
- Narayan et al. (2019). Noncontact identification of sleep-disturbed breathing from smartphone-recorded sounds validated by polysomnography. https://doi.org/10.1007/s11325-018-1695-6
- Sommer et al. (2024). Evaluation of an OSA Risk Screening Smartphone App in a General, Non-Symptomatic Population Sample (ESOSA). https://doi.org/10.3390/jcm13164664
- Han et al. (2024). In-Home Smartphone-Based Prediction of Obstructive Sleep Apnea in Conjunction With Level 2 Home Polysomnography. https://doi.org/10.1001/jamaoto.2023.3490
- Cho et al. (2022). Evaluating Prediction Models of Sleep Apnea From Smartphone-Recorded Sleep Breathing Sounds. https://doi.org/10.1001/jamaoto.2022.0244
- Romero et al. (2022). Acoustic Screening for Obstructive Sleep Apnea in Home Environments Based on Deep Neural Networks. https://doi.org/10.1109/JBHI.2022.3154719
- Le et al. (2023). Real-Time Detection of Sleep Apnea Based on Breathing Sounds and Prediction Reinforcement Using Home Noises: Algorithm Development and Validation. https://doi.org/10.2196/44818
- Stephens et al. (2026). Evaluation of a sleep sound analysis smartphone app for home obstructive sleep apnea screening among community-dwelling Chinese adults with hypertension. https://doi.org/10.1007/s11325-026-03754-4
- Frija et al. (2025). Proposition of a new, minimally-invasive, software smartphone device to predict sleep apnea and its severity. https://doi.org/10.1007/s11325-025-03441-w
- Zhou et al. (2025). Pioneering Contactless Sleep Apnea Detection: Harnessing the Power of Smartphones. https://doi.org/10.1109/EMBC58623.2025.11252641
- Edouard et al. (2021). Validation of the Withings Sleep Analyzer, an under-the-mattress device for the detection of moderate-severe sleep apnea syndrome. https://doi.org/10.5664/jcsm.9168
- Lechat et al. (2026). Zero burden multi night monitoring with AI enabled technology reduces obstructive sleep apnea misdiagnosis. https://doi.org/10.1038/s41746-026-02914-w
- Jagielski et al. (2023). Evaluating an under-mattress sleep monitor compared to a peripheral arterial tonometry home sleep apnea test device in the diagnosis of obstructive sleep apnea. https://doi.org/10.1007/s11325-022-02751-7
- Ding et al. (2022). Polysomnographic validation of an under-mattress monitoring device in estimating sleep architecture and obstructive sleep apnea in adults. https://doi.org/10.1016/j.sleep.2022.04.010
- Hasanain et al. (2025). Diagnostic value of wearable and contactless technologies for detecting obstructive sleep apnea: a systematic review and meta-analysis. https://doi.org/10.1007/s11325-025-03544-4
- Khalil et al. (2023). Validity of non-contact methods for diagnosis of Obstructive Sleep Apnea: a systematic review and meta-analysis. https://doi.org/10.1016/j.jclinane.2023.111087
- Kim et al. (2022). Diagnostic value of smartphone in obstructive sleep apnea syndrome: A systematic review and meta-analysis. https://doi.org/10.1371/journal.pone.0268585
- Tan et al. (2025). Machine Listening for OSA Diagnosis: A Bayesian Meta-Analysis. https://doi.org/10.1016/j.chest.2025.04.006
- da Silva Dantas et al. (2025). Oximetry-based devices in diagnosis of obstructive sleep apnea: A systematic review and meta-analysis. https://doi.org/10.1016/j.smrv.2025.102139
- Jin et al. (2026). Performance evaluation of finger-worn devices for sleep stage classification and sleep apnea detection: a systematic review and meta-analysis. https://doi.org/10.1186/s12967-026-08214-x
- Abd-Alrazaq et al. (2024). Detection of Sleep Apnea Using Wearable AI: Systematic Review and Meta-Analysis. https://doi.org/10.2196/58187
- Duarte et al. (2023). The Role of Novel Digital Clinical Tools in the Screening or Diagnosis of Obstructive Sleep Apnea: Systematic Review. https://doi.org/10.2196/47735
- Baptista et al. (2022). A systematic review of smartphone applications and devices for obstructive sleep apnea. https://doi.org/10.1016/j.bjorl.2022.01.004
- Pillion et al. (2026). Accuracy of artificial intelligence-based prediction of moderate-to-severe obstructive sleep apnea using audio signals: A systematic review. https://doi.org/10.1016/j.smrv.2026.102357
- US Food and Drug Administration (2023). 510(k) K231667 (under-mattress sleep apnea device; clinical data from Edouard et al. 2021). https://www.accessdata.fda.gov/cdrh_docs/pdf23/K231667.pdf
Corrections
None yet.
Found an error? Email hello@paloaltohealthlabs.com.
More from the Palo Alto Health Labs Co-Scientist · All posts
Have your own questions? Sign up for the Beta
Tell us the health AI or digital health question you would want the Co-Scientist to review, and join the beta list.
Scope it in 30 minutes
Book a scoping call and we'll name the smallest engagement that moves you forward. Most teams start with the fixed-scope Initial Audit.
Book a 30-min scoping callPrefer to write? Send a short note.