The short version
- An artificial-intelligence reading of a standard 12-lead ECG separates people with a weak heart pump from people without one quite well. Across 24 separate patient groups that were not used to build the models, the pooled area under the curve (AUC) was 0.912 (95% CI 0.896–0.926), with low ejection fraction (EF) defined as 40% or below.
- Results vary by setting. The 95% prediction interval, the range expected for a new hospital, runs from 0.817 to 0.960.
- Studies that used a 35% cut-off are kept separate. There are only 4 such groups, and they disagree with each other (pooled 0.868, 95% CI 0.640–0.961).
- Single-lead and wearable recordings (smartwatches, ECG-enabled stethoscopes, handheld devices) did somewhat less well: pooled AUC 0.875 (95% CI 0.781–0.932) across 6 studies.
- Good AUC does not mean most positive results are correct. When few people have the condition, most positives are false alarms (see "What a positive result means").
- Four randomised trials tested what happens when clinicians receive the AI result. Their outcomes are reported on their own and are not mixed with the accuracy figures.
Why this matters
Ejection fraction is the share of blood the main pumping chamber pushes out with each beat. A low EF can go unnoticed for years and is treatable once found. An echocardiogram measures it directly but needs equipment and trained staff. An ECG is cheap and everywhere. The question here is how well AI reading of an ECG finds low EF when tested on new patients.
How the evidence was gathered
-
Search: PubMed only (NCBI E-utilities esearch), run on 23 November 2026 with 4 query sets, plus a top-200 OpenAlex check. Date window: publication dates 2018 to 2026 (
2018:2026[dp], which PubMed translates as 2018/01/01 to 2026/12/31) for the two accuracy-study queries; the review and trial queries had no date limit. The exact PubMed strings, copied from search/search_log.json, were:- Q1 (primary):
("artificial intelligence"[tiab] OR "deep learning"[tiab] OR "machine learning"[tiab] OR "neural network"[tiab] OR "AI-ECG"[tiab] OR "AI-enabled"[tiab]) AND (electrocardiogra*[tiab] OR ECG[tiab] OR EKG[tiab]) AND ("ejection fraction"[tiab] OR "ventricular systolic dysfunction"[tiab] OR "ventricular dysfunction"[tiab] OR "reduced ejection"[tiab] OR HFrEF[tiab]) AND 2018:2026[dp] - Q2 (broadened):
("artificial intelligence"[tiab] OR AI[tiab] OR "deep learning"[tiab] OR "deep neural"[tiab] OR "convolutional"[tiab] OR "machine learning"[tiab]) AND (electrocardiogra*[tiab] OR ECG[tiab] OR ECGs[tiab] OR EKG[tiab]) AND (LVEF[tiab] OR LVSD[tiab] OR "systolic dysfunction"[tiab] OR "low ejection"[tiab] OR "reduced ejection"[tiab] OR "ejection fraction"[tiab] OR "contractile dysfunction"[tiab]) AND 2018:2026[dp] - Q3 (reviews meta):
("artificial intelligence"[tiab] OR "deep learning"[tiab] OR "machine learning"[tiab]) AND (electrocardiogra*[tiab] OR ECG[tiab]) AND ("ejection fraction"[tiab] OR "systolic dysfunction"[tiab]) AND (meta-analysis[pt] OR systematic review[pt] OR "meta-analysis"[tiab] OR "systematic review"[tiab]) - Q4 (trials):
("artificial intelligence"[tiab] OR "deep learning"[tiab] OR "AI-enabled"[tiab] OR "AI-ECG"[tiab]) AND (electrocardiogra*[tiab] OR ECG[tiab]) AND ("ejection fraction"[tiab] OR "systolic dysfunction"[tiab] OR cardiomyopath*[tiab]) AND (randomized controlled trial[pt] OR random*[tiab] OR pragmatic[tiab] OR prospective[tiab]) - OpenAlex check:
artificial intelligence electrocardiogram low ejection fraction left ventricular systolic dysfunction, filtered to publication dates 2018-01-01 to 2026-12-31; the top 200 results by relevance were examined. 577 records were screened and 56 were kept: 41 accuracy studies, 4 randomised trials and 11 earlier reviews (context only). Embase, Scopus and trial registries were not searched.
- Q1 (primary):
-
What counted: adults; AI applied to an ECG; EF measured by echocardiography or cardiac MRI; and accuracy reported in patients outside the model's own development data (external, prospective or later-in-time testing). One study used a 50% cut-off and was dropped at full text, leaving 40 accuracy studies for synthesis.
- Who did the work: screening and data extraction were done by an AI assistant as a single reviewer. Every extracted number was matched automatically against the source abstract or full text. Each paper's DOI was re-checked against its PubMed record and the Crossref title.
- Pooling: AUCs were combined on the logit scale with a random-effects model (REML estimator), with the DerSimonian–Laird method as a check. The main 95% CI uses the Hartung–Knapp method with a t distribution on k−1 degrees of freedom. It allows for the uncertainty in how much studies differ, and is wider when few studies are pooled. The Hartung–Knapp scaling factor q is floored at 1, so this interval is never narrower than the Wald interval. In formula terms, the standard error is the Wald standard error × √max(1, q). The scaling factor q was 1.000 for the 12-lead 40% pool, 0.985 for the 35% pool and 0.931 for the wearable pool. Values for every sensitivity analysis are in analysis/pool_results_v4.md. The usual normal-based (Wald) 95% CI is reported alongside. The 95% prediction interval uses a t distribution with k−2 degrees of freedom and the REML variance tau² + SE², where SE is the unadjusted (Wald) standard error of the pooled mean. The Hartung–Knapp floor is not applied to the prediction interval. Tau is the standard deviation of the true AUCs between settings on the logit scale. Multiplying and dividing the odds AUC/(1−AUC) of the average setting by exp(1.96 × tau) spans about 95% of settings. Standard errors came from each study's 95% CI, including Attia 2021 0.82 (0.75–0.90), Jeong 2024 0.78 (0.71–0.85) and Park YS 2026 0.901 (0.865–0.937), whose CIs are printed in the full texts but not the abstracts. Each patient group enters a pool only once.
- Software: the pooling is custom Python code (analysis/pool_auc_v4.py), run with Python 3.13.5, numpy 2.5.5, scipy 1.18.1 and statsmodels 0.15.0 (statsmodels for the DerSimonian–Laird check). Figures were drawn with matplotlib 3.11.2. A separate REML fit (analysis/pi_check_v4.py) reproduced every prediction interval to 3 decimals.
Avoiding double counting
Some papers re-test the same patients, so each group was counted once:
- Chagas disease cohort: two papers report the same 1,304 Brazilian participants (Brito 2021; Cardoso 2026). Only the earlier test of the unmodified model (Brito 2021) is pooled. Swapping in the later, fine-tuned model barely moves the result (0.914 instead of 0.912).
- Other overlaps: one study reused part of the Brazilian ELSA-Brasil cohort, one Korean hospital network appeared in more than one paper, and one study tested 4 models on the same 1,203 patients. In each case one row was kept and the alternatives were tested in sensitivity analyses. These rows appear as open circles in Figures 1 and 3.
- Independence of the test data: five studies applied an existing commercial model in new hospitals. Their full texts were checked to see whether the test patients were truly separate from the development data. Three said so explicitly and were pooled. Two did not, and were left out of the pool.
Results
1. Standard 12-lead ECG, EF 40% or below

Figure 1. Pool A: standard 12-lead ECG, EF 40% or below. Each row is one patient group. Filled blue squares are pooled. Open grey circles in the grey band are shown for comparison but not pooled, because they reuse the same patients or for another stated reason. The diamond is the pooled estimate; its width is the floored Hartung–Knapp 95% CI. The bar beneath it is the prediction interval.
- Size of the evidence: 24 patient groups from 19 studies, covering emergency departments, hospital echo labs, health screening, pregnancy, people with sepsis, people with left bundle branch block and people with Chagas disease, in North and South America, Europe, Africa and Asia.
- Range: individual AUCs ran from 0.780, in people with atrial fibrillation and a fast heart rate (Jeong 2024), to 0.973 in a low-risk screening population (Rhee 2026).
- Pooled AUC: 0.912 (Hartung–Knapp 95% CI 0.896–0.926; Wald 95% CI 0.897–0.925; prediction interval 0.817–0.960). Tau is 0.397 on the logit scale, so the odds AUC/(1−AUC) in a typical new setting could be about 2.18 times higher or lower than the average.
- Sensitivity checks: the result held across every check: counting each study once, 0.913; using the weakest or strongest of the 4 models tested head-to-head, 0.910 or 0.915; and dropping a study that sampled cases and controls 50/50, 0.911.
- Heterogeneity: very high (I² 97%). This is common when many large, precise studies are combined. The prediction interval is the more useful guide.
- Cut-off choice: in some studies the cut-off for calling a result positive was chosen in the same data (Koenig 2024; Park 2026). Their sensitivity and specificity are likely to look better than they would in new patients, though their AUCs are not affected.
2. Standard 12-lead ECG, EF 35% or below

Figure 2. Pool B: standard 12-lead ECG, EF 35% or below, pooled separately from pool A. Symbols as in Figure 1. The arrow shows that the prediction interval runs past the left edge of the axis.
- Groups: four groups from 3 studies: a held-out New York hospital (Vaid 2022, 0.95); a US academic centre (Hughes 2024, Columbia, 0.88); the UK Biobank general population (Hughes 2024, 0.74); and a Russian population sample in which only 26 of 4,277 people had low EF (Attia 2021, 0.82, 95% CI 0.75–0.90).
- Pooled AUC: 0.868 (Hartung–Knapp 95% CI 0.640–0.961; Wald 95% CI 0.746–0.937; tau 0.811). The prediction interval (0.117–0.997) is too wide to be informative.
- What drives the spread: the general-population groups did worse than hospital groups. With 4 groups this is suggestive only.
- A sensitivity analysis combining both cut-off families (one row per group) gave 0.904, but this mixes definitions and is not the main result.
3. Single-lead and wearable recordings, EF 40% or below

Figure 3. Pool C: single-lead or wearable recordings, EF 40% or below. Symbols as in Figure 1. The grey band holds one analysis that took a single lead from a 12-lead ECG, shown for comparison only. The arrow shows that the prediction interval runs past the left edge of the axis.
- Studies: six studies: two ECG-enabled stethoscope studies (Bachtiger 2022, 0.85; Prakash Borde 2026, 0.73); two smartwatch studies (Attia 2022, 0.885; Kwon 2022, 0.934); a 6-lead handheld device (Lim 2025, 0.924); and a one-lead handheld recorder (Attia 2022, 0.85, with only 8 people with low EF).
- Pooled AUC: 0.875 (Hartung–Knapp 95% CI 0.781–0.932; Wald 95% CI 0.807–0.921; prediction interval 0.531–0.977; tau 0.601).
- Excluded data: some studies removed poor-quality recordings before analysis (for example Prakash Borde 2026). Real-world performance could therefore be lower.
- Not counted as wearables: analyses that simply took one lead out of a hospital 12-lead ECG (for example Cho 2021) are shown for comparison only.
What a positive result means
AUC describes how well a test ranks people. It does not tell a patient how likely a positive result is to be right. That depends on how common low EF is in the group being tested:
- Hospital or high-risk patients: in the Chagas cohort, 7.1% had low EF and 25% of positives were confirmed (Brito 2021).
- Middle-aged community cohort: 1.13% had low EF, and 24.4% of positives were confirmed (Santana Junior 2025).
- Low-risk screening: only 0.054% had low EF (32 of 60,711 ECG–echo pairs). Even with 99.4% specificity, 7.7% of positives were confirmed, although almost no cases were missed (negative predictive value 100%; Rhee 2026).
So a positive AI-ECG is best read as "arrange an echocardiogram", not as a diagnosis. A negative result in a low-risk person is reassuring.
What happens in practice: the randomised trials
These trials measured new diagnoses in routine care, not test accuracy. Three of them defined the outcome as EF 50% or below (or below 50%), not the 40% used above:
- US primary care (cluster trial; EAGLE, Yao 2021): new low-EF diagnoses within 90 days were 2.1% with the AI result versus 1.6% without (odds ratio 1.32, 95% CI 1.01–1.61), with little change in echocardiogram use (19.2% vs 18.2%).
- Taiwan, hospital inpatients under non-cardiologists (Tsai 2025): new diagnoses within 30 days were 1.5% versus 1.1% (hazard ratio 1.50, 95% CI 1.11–2.03).
- Nigeria, pregnant and postpartum women (SPEC-AI, Adedinsewo 2024): an AI-enabled stethoscope doubled detection, 4.1% versus 2.0% (odds ratio 2.12, 95% CI 1.05–4.27). The 12-lead AI-ECG arm was not clearly different (odds ratio 1.75, 95% CI 0.85–3.62).
- UK primary care, 205 practices (TRICORDER, Kelshiker 2026): an AI stethoscope made no clear difference to new heart failure diagnoses (rate ratio 0.94, 95% CI 0.86–1.02). Intervention practices, with 701,933 registered patients, recorded 12,725 patient examinations with the AI stethoscope across 972 clinical users. The trial authors report that use of the AI stethoscope was associated with higher detection rates.
Together these trials suggest the result can raise diagnosis rates when clinicians actually use it. None measured whether patients lived longer or felt better.
Who ran the studies
- The pattern: many studies involve the model's developers, in different ways. In some, company staff were authors, the company funded the work or provided the software, or the developer institution held a licensing interest. In others, the academic team that built the model tested it in new hospitals. In a minority, researchers with no stated link to the developer did the testing.
- The 24 pooled 12-lead groups:
- 10 came from studies declaring commercial or licensing involvement;
- 10 came from academic developer teams testing their own models;
- 3 were third-party tests: Koenig 2024 and Lee MJ 2026 (no developer role declared beyond technical help running the software), and Croon 2026, a head-to-head test of 4 shared models;
- 1 did not state the developer's role.
- The subsets compared: pooled AUCs were similar: 0.921 for the groups with commercial or licensing involvement and 0.910 for the rest. Third-party tests alone gave 0.904, but rest on only 3 groups.
- The trials: the developer institution or a licensee was involved in some of them (for example, it funded SPEC-AI). The Taiwan trial (Tsai 2025) tested a model built at the trial hospital; it was funded by a national science council, and its authors declared no competing interests.
- What the comparison can and cannot say: these facts are reported so readers can weigh them. They do not show that involvement raised or lowered results; a comparison this small and this confounded by setting cannot answer that. What can be said is that third-party tests exist and give results in the same range.
Limits of this review
- Scope: one database, a single AI reviewer, and a non-systematic search. Some relevant studies are probably missing, especially non-English and non-PubMed work.
- Units: studies counted patients, ECG–echo pairs or emergency visits. These were pooled together.
- Overlap: some US multi-site datasets may share patients, which could not be checked from the papers.
- No formal checks: no formal risk-of-bias scoring or publication-bias test was done.
- Thin strata: the 35% and wearable pools rest on few studies.
Bottom line
On a standard ECG, AI separates people with and without EF 40% or below well in new hospitals: a typical AUC around 0.91, with an expected range of about 0.82–0.96 in a new setting. Wearable and single-lead versions do somewhat less well and vary more. In low-risk groups most positives will be false alarms, so a positive result should lead to an echocardiogram, not a diagnosis. Trials show more cases found when clinicians act on the result, but no trial yet shows better patient outcomes.
References
Studies named in the text: Attia 2021; Attia 2022 (smartwatch); Attia 2022 (one-lead); Bachtiger 2022; Brito 2021; Cardoso 2026; Cho 2021; Croon 2026; Hughes 2024; Jeong 2024; Kelshiker 2026; Koenig 2024; Kwon 2022; Lee MJ 2026; Lim 2025; Park 2026; Prakash Borde 2026; Rhee 2026; Santana Junior 2025; Tsai 2025; Vaid 2022; Yao 2021; Adedinsewo 2024.
- Adedinsewo D, Johnson PW, Morales-Lara C, et al. Validating ECG-Based Artificial Intelligence Models for Cardiomyopathy Detection in Obstetric Patients. JACC: Advances. 2026. https://doi.org/10.1016/j.jacadv.2026.103321
- Adedinsewo DA, Morales-Lara AC, Afolabi BB, et al. Artificial intelligence guided screening for cardiomyopathies in an obstetric population: a pragmatic randomized clinical trial. Nat Med. 2024. https://doi.org/10.1038/s41591-024-03243-9
- Attia IZ, Tseng AS, Benavente ED, et al. External validation of a deep learning electrocardiogram algorithm to detect ventricular dysfunction. International Journal of Cardiology. 2021. https://doi.org/10.1016/j.ijcard.2020.12.065
- Attia ZI, Harmon DM, Dugan J, et al. Prospective evaluation of smartwatch-enabled detection of left ventricular dysfunction. Nat Med. 2022. https://doi.org/10.1038/s41591-022-02053-1
- Attia ZI, Dugan J, Rideout A, et al. Automated detection of low ejection fraction from a one-lead electrocardiogram: application of an AI algorithm to an electrocardiogram-enabled Digital Stethoscope. European Heart Journal - Digital Health. 2022. https://doi.org/10.1093/ehjdh/ztac030
- Bachtiger P, Petri CF, Scott FE, et al. Point-of-care screening for heart failure with reduced ejection fraction using artificial intelligence during ECG-enabled stethoscope examination in London, UK: a prospective, observational, multicentre study. The Lancet Digital Health. 2022. https://doi.org/10.1016/s2589-7500(21)00256-9
- Brito BODF, Attia ZI, Martins LNA, et al. Left ventricular systolic dysfunction predicted by artificial intelligence using the electrocardiogram in Chagas disease patients–The SaMi-Trop cohort. PLoS Negl Trop Dis. 2021. https://doi.org/10.1371/journal.pntd.0009974
- Cardoso CS, Mangold K, Padilha da Silva JL, et al. AI-ECG for Detecting Left Ventricular Systolic Dysfunction in Chagas Disease. JACC: Advances. 2026. https://doi.org/10.1016/j.jacadv.2026.102878
- Carter RE, Johnson PW, Strom JB, et al. Multisite, External Validation of an AI-Enabled ECG Algorithm for Detection of Low Ejection Fraction. JACC: Advances. 2026. https://doi.org/10.1016/j.jacadv.2025.102537
- Cho J, Lee B, Kwon JM, et al. Artificial Intelligence Algorithm for Screening Heart Failure with Reduced Ejection Fraction Using Electrocardiography. ASAIO Journal. 2021. https://doi.org/10.1097/mat.0000000000001218
- Croon PM, Boonstra MJ, Allaart CP, et al. Artificial Intelligence–Enhanced Electrocardiogram Models for Detection of Left Ventricular Dysfunction. JACC: Advances. 2026. https://doi.org/10.1016/j.jacadv.2025.102572
- Hughes JW, Somani S, Elias P, et al. Simple models vs. deep learning in detecting low ejection fraction from the electrocardiogram. European Heart Journal - Digital Health. 2024. https://doi.org/10.1093/ehjdh/ztae034
- Jeong JH, Kang S, Lee HS, et al. Deep learning algorithm for predicting left ventricular systolic dysfunction in atrial fibrillation with rapid ventricular response. European Heart Journal - Digital Health. 2024. https://doi.org/10.1093/ehjdh/ztae062
- Kelshiker MA, Bächtiger P, Petri CF, et al. Triple cardiovascular disease detection with an artificial intelligence-enabled stethoscope (TRICORDER) in the UK: a cluster-randomised controlled implementation trial. The Lancet. 2026. https://doi.org/10.1016/s0140-6736(25)02156-7
- König S, Hohenstein S, Nitsche A, et al. Artificial intelligence-based identification of left ventricular systolic dysfunction from 12-lead electrocardiograms: external validation and advanced application of an existing model. European Heart Journal - Digital Health. 2024. https://doi.org/10.1093/ehjdh/ztad081
- Kwon JM, Kim KH, Jeon KH, et al. Development and Validation of Deep-Learning Algorithm for Electrocardiography-Based Heart Failure Identification. Korean Circ J. 2019. https://doi.org/10.4070/kcj.2018.0446
- Kwon JM, Jo YY, Lee SY, et al. Artificial Intelligence-Enhanced Smartwatch ECG for Heart Failure-Reduced Ejection Fraction Detection by Generating 12-Lead ECG. Diagnostics. 2022. https://doi.org/10.3390/diagnostics12030654
- Lee H, Kwon WY, Song KJ, et al. Interethnic validation of electrocardiogram image analysis software for detecting left ventricular dysfunction in an emergency department population. Clin Exp Emerg Med. 2025. https://doi.org/10.15441/ceem.24.342
- Lee H, Kim Y, Sykora D, et al. External validation of ECG artificial intelligence for emergency and cardiac assessment across a large-scale U.S. healthcare system. npj Digit. Med. 2026. https://doi.org/10.1038/s41746-026-02682-7
- Lee MJ, Jung H. Prediction of Left Ventricular Systolic Dysfunction Using an Artificial Intelligence-Based Electrocardiogram Analysis Model in Patients Presenting to the Emergency Department. Diagnostics. 2026. https://doi.org/10.3390/diagnostics16162587
- Lee S, Son JW, Kim SA, et al. Deep Learning Model Using Transfer Learning for Detecting Left Ventricular Systolic Dysfunction: Retrospective Algorithm Development and Validation Study. JMIR Med Inform. 2026. https://doi.org/10.2196/83127
- Lee SY, Yoo AH, Kang S, et al. Detection and prognostic stratification of left ventricular systolic dysfunction in left bundle branch block using an artificial intelligence–enabled electrocardiography. J Cardiovasc Imaging. 2026. https://doi.org/10.1186/s44348-026-00066-9
- Lim J, Lee HS, Han GI, et al. Artificial intelligence-enhanced six-lead portable electrocardiogram device for detecting left ventricular systolic dysfunction: a prospective single-centre cohort study. European Heart Journal - Digital Health. 2025. https://doi.org/10.1093/ehjdh/ztaf025
- Pandey A, Keshvani N, Segar MW, et al. Artificial Intelligence Electrocardiogram and Left Ventricular Systolic Dysfunction in Kenya. JAMA Cardiol. 2026. https://doi.org/10.1001/jamacardio.2026.0908
- Park YS, Kim J, Jo YH, et al. Screening for Left Ventricular Dysfunction in Sepsis Using a Smartphone ECG Analysis Application: A Multicenter Validation Study. J Korean Med Sci. 2026. https://doi.org/10.3346/jkms.2026.41.e217
- Prakash Borde D, Joshi S, Chidambaram K, et al. Digital Stethoscope–Derived Single-Lead Electrocardiogram and Artificial Intelligence to Detect Low Ejection Fraction: A Multicenter Observational Study. Journal of Cardiothoracic and Vascular Anesthesia. 2026. https://doi.org/10.1053/j.jvca.2026.03.002
- Rhee TM, Kang S, Lee MS, et al. Artificial Intelligence–Driven Electrocardiogram Screening for Asymptomatic Left Ventricular Systolic Dysfunction in the General Population. JACC: Advances. 2026. https://doi.org/10.1016/j.jacadv.2026.102660
- Sangha V, Nargesi AA, Dhingra LS, et al. Detection of Left Ventricular Systolic Dysfunction From Electrocardiographic Images. Circulation. 2023. https://doi.org/10.1161/circulationaha.122.062646
- Santana Júnior WBD, Pinto Filho MM, Barreto SM, et al. Uso da Inteligência Artificial Aplicada ao Eletrocardiograma para Diagnóstico de Disfunção Sistólica Ventricular Esquerda. Arq. Bras. Cardiol. 2025. https://doi.org/10.36660/abc.20240740
- Tsai DJ, Lin C, Liu WT, et al. Artificial intelligence-assisted diagnosis and prognostication in low ejection fraction using electrocardiograms in inpatient department: a pragmatic randomized controlled trial. BMC Med. 2025. https://doi.org/10.1186/s12916-025-04190-z
- Vaid A, Johnson KW, Badgeley MA, et al. Using Deep-Learning Algorithms to Simultaneously Identify Right and Left Ventricular Dysfunction From the Electrocardiogram. JACC: Cardiovascular Imaging. 2022. https://doi.org/10.1016/j.jcmg.2021.08.004
- Yagi R, Goto S, Katsumata Y, et al. Importance of external validation and subgroup analysis of artificial intelligence in the detection of low ejection fraction from electrocardiograms. European Heart Journal - Digital Health. 2022. https://doi.org/10.1093/ehjdh/ztac065
- Yao X, Rushlow DR, Inselman JW, et al. Artificial intelligence–enabled electrocardiograms for identification of patients with low ejection fraction: a pragmatic, randomized clinical trial. Nat Med. 2021. https://doi.org/10.1038/s41591-021-01335-4
Corrections
None yet.
Found an error? Email hello@paloaltohealthlabs.com.
More from the Palo Alto Health Labs Co-Scientist · All posts
Have your own questions? Sign up for the Beta
Tell us the health AI or digital health question you would want the Co-Scientist to review, and join the beta list.
Scope it in 30 minutes
Book a scoping call and we'll name the smallest engagement that moves you forward. Most teams start with the fixed-scope Initial Audit.
Book a 30-min scoping callPrefer to write? Send a short note.