Prepared with the Palo Alto Health Labs Co-Scientist. Screening and data extraction were done by AI. Searched 23 November 2026. Version 2b, revised 24 November 2026.
Key message: Across 40 randomized trials, colonoscopy with real-time AI polyp detection found at least one adenoma (a polyp that can turn into cancer) in about 16% more patients than colonoscopy without it (risk ratio 1.16, 95% CI 1.12 to 1.20). In absolute terms, that is about 6.5 more patients with an adenoma found per 100 colonoscopies. The gain was smaller in newer trials and where doctors already found adenomas often without AI. The extra finds are mostly small adenomas, and no pooled trial measured whether fewer people get or die of bowel cancer.
Results
The overall effect is modest but consistent. The adenoma detection rate (ADR) is the share of patients in whom at least one adenoma is found. It is the standard quality measure for colonoscopy. Pooling 40 trials (32,037 patients) gave a risk ratio of 1.16 (Hartung-Knapp 95% CI 1.12 to 1.20; Wald 95% CI 1.12 to 1.20). In 38 of the 40 trials, more patients had an adenoma found in the AI arm; in 2, fewer did.
In absolute terms: about 6.5 more patients per 100. Across the 36 trials that report patient counts for both arms, AI added 6.5 percentage points to the ADR (95% CI 5.1 to 7.9). The 95% prediction interval was 1.3 to 11.7 points. As an illustration only: at the median ADR without AI in these trials (41.9%), a risk ratio of 1.16 would mean about 48.5% with AI.
The trials vary moderately. The spread between trials was moderate (tau 0.067 and tau² 0.0044 on the log scale; I² 44%). The 95% prediction interval for all 40 trials, the range where a new trial in a similar setting would likely land, was 1.01 to 1.33. That means a gain in most similar settings, though sometimes close to none. Dropping any single trial moved the pooled risk ratio only between 1.15 and 1.17.
Newer trials and high-detection settings show smaller gains. Trials published up to 2022 (8 trials) pooled to 1.24 (95% CI 1.14 to 1.34). Trials from 2023 on (32 trials) pooled to 1.14 (95% CI 1.10 to 1.18); their own prediction interval was 0.99 to 1.31, which includes no effect. Where the ADR without AI was below 40%, the pooled risk ratio was 1.20 (17 trials); where it was 40% or higher, it was 1.14 (23 trials). Newer trials also had higher ADRs without AI (median 43.2% from 2023 on vs 34.2% up to 2022), so these two patterns overlap. Both are comparisons between trials, so they show a pattern, not a cause.
Baseline detection rate (exploratory, not an established effect modifier). The gain from AI was smaller in trials where endoscopists without AI already found adenomas in more patients (Figure 4). In an exploratory meta-regression of the 40 trials, each 10-point higher ADR in the control arm went with a risk ratio about 4.6% lower (ratio of risk ratios 0.954; slope −0.047 on the log risk-ratio scale, 95% CI −0.071 to −0.023; p < 0.001). In this model, baseline ADR accounted for about 64% of the variation between trials. The predicted risk ratio was about 1.24 where the control arm found adenomas in 30% of patients, 1.19 at 40%, 1.13 at 50% and 1.08 at 60%. Part of this slope may be regression to the mean. The control-arm rate appears in both the covariate and the risk ratio, so a control arm that is low by chance raises the risk ratio and lowers the covariate at the same time. Using the average of both arms' rates instead gave a flatter slope (−0.032 per 10 points, 95% CI −0.061 to −0.002, p = 0.036; 36 trials). That slope still points down, but it is about a third smaller. Once baseline ADR was in the model, publication year no longer explained the smaller gains in newer trials. The year p-value was 0.33 for a 2023-or-later term and 0.26 for year as a number, each added to baseline ADR alone. It was 0.29 when centre type was also added (single-centre term p = 0.38). On its own, a 2023-or-later term had p = 0.047. This analysis was added after peer review, was not planned in advance and compares trials, not patients. It suggests a pattern, but it does not show that baseline ADR changes how well AI works.
Small trials and trial design. Smaller trials and single-centre trials tended to show larger gains from AI (single-centre trials: risk ratio 1.20, 95% CI 1.12 to 1.29, prediction interval 0.97 to 1.48, 17 trials), but the overall result did not depend on them. Restricting to the 23 multicentre trials gave 1.14 (95% CI 1.10 to 1.19; prediction interval 1.02 to 1.28). Leaving out the 4 single-centre trials from mainland China gave 1.15 (95% CI 1.11 to 1.19). Also leaving out the single-centre trial from Hong Kong gave 1.14 (95% CI 1.11 to 1.18). Dropping the smallest trial, the 3 smallest trials, or every trial smaller than the median size (800 patients) left the pooled risk ratio between 1.15 and 1.16. A funnel plot showed mild asymmetry (Figure 5; Egger test p = 0.10; rank correlation test, Kendall's tau 0.17, p = 0.12). A trim-and-fill adjustment (L0 method; descriptive only) added 6 hypothetical small trials with no gain and gave a risk ratio of 1.14 (95% CI 1.10 to 1.18); the R0 method added none. These checks have limited power. The asymmetry may reflect real differences between trials, such as baseline ADR, rather than unpublished studies.
The result holds across checks. Using only trials with arm-level counts (35 trials) gave 1.16 (95% CI 1.11 to 1.20). Leaving out trials where both arms also used an add-on, such as an imaging mode, a cuff device or water exchange (36 trials left), gave 1.16 (95% CI 1.12 to 1.20). The one sham-controlled trial stays in this check, because a sham system is a way of hiding the arm, not an add-on. Trials whose authors declared industry ties pooled to 1.15 (13 trials). Trials declaring none pooled to 1.18 (10 trials), and trials with no statement in the PubMed record to 1.16 (17 trials).
Trials that could not be pooled point the same way. 13 more trials met the criteria, but their fetched source lacked arm sizes or ADR numbers, or had internal inconsistencies. Of the 10 that report ADR for both arms, 9 show a higher rate with AI and 1 a lower one.
Studies of routine practice are mixed. The first three studies below were left out of the pool because they did not randomize individual patients, but they show how AI performs in routine care:
- A trial in a Spanish surveillance programme randomized colonoscopy days rather than patients. It found no overall difference (ADR 58.8% with AI vs 57.4% without; adjusted risk ratio 1.02, 95% CI 0.95 to 1.10), with a gain only among endoscopists with lower ADRs [54].
- A US Veterans Affairs study randomized whole facilities to receive AI devices. Across about 334,000 colonoscopies, ADR rose 4.2 percentage points at sites with AI and fell 0.7 points at control sites (odds ratio 1.22, 95% CI 1.15 to 1.28) [55].
- A US pragmatic implementation study, which used historical and concurrent controls rather than randomization, found that AI did not improve colonoscopists' detection measures [56].
- Among the pooled trials, the US community-centre trial AI-SEE found no difference in ADR (35.9% vs 37.2%) [36].
How this fits with other evidence
Earlier reviews found a similar or slightly larger gain. A 2024 meta-analysis of 44 randomized trials, searched to February 2024, found a risk ratio of 1.21 (95% CI 1.15 to 1.28) for ADR, with no difference in advanced neoplasia per colonoscopy [59]. A 2025 review of 28 trials found 1.20 (95% CI 1.14 to 1.27), mostly from more small adenomas and with no clear gain in advanced adenomas [60]. The American Gastroenterological Association (AGA) 2025 guideline reported an 8% increase in ADR (95% CI 6% to 10%) [58]. Most trials here overlap with those reviews. This review adds trials from 2025 and 2026 and leaves out back-to-back (tandem) designs. Its somewhat smaller estimate fits with newer trials in settings where ADR is already high.
More adenomas found does not yet mean fewer cancers. The AGA guideline made no recommendation for or against AI detection, because the certainty of evidence for the outcomes that matter most was very low [58]. Its modelling estimated 11 fewer bowel cancers and 2 fewer bowel cancer deaths per 10,000 people, against 635 more surveillance colonoscopies [58]. No pooled trial measured cancers found between colonoscopies or cancer deaths.
Figures





Summary table
| Analysis | Trials | Risk ratio | Hartung-Knapp 95% CI | Wald 95% CI | tau | tau² | I² | q | 95% prediction interval |
|---|---|---|---|---|---|---|---|---|---|
| All pooled trials | 40 | 1.16 | 1.12 to 1.20 | 1.12 to 1.20 | 0.067 | 0.0044 | 44% | 0.98 | 1.01 to 1.33 |
| Arm-level counts only | 35 | 1.16 | 1.11 to 1.20 | 1.12 to 1.20 | 0.067 | 0.0045 | 44% | 1.00 | 1.00 to 1.33 |
| No shared add-on in both arms | 36 | 1.16 | 1.12 to 1.20 | 1.12 to 1.20 | 0.068 | 0.0046 | 45% | 0.97 | 1.00 to 1.34 |
| Published 2023 or later | 32 | 1.14 | 1.10 to 1.18 | 1.10 to 1.18 | 0.065 | 0.0043 | 44% | 0.95 | 0.99 to 1.31 |
| Published 2022 or earlier | 8 | 1.24 | 1.14 to 1.34 | 1.15 to 1.33 | 0.052 | 0.0027 | 21% | 0.87 | 1.06 to 1.44 |
| ADR without AI below 40% | 17 | 1.20 | 1.12 to 1.29 | 1.13 to 1.28 | 0.086 | 0.0074 | 41% | 0.93 | 0.99 to 1.46 |
| ADR without AI 40% or more | 23 | 1.14 | 1.10 to 1.18 | 1.10 to 1.18 | 0.055 | 0.0030 | 41% | 0.97 | 1.01 to 1.28 |
| Industry ties declared | 13 | 1.15 | 1.10 to 1.20 | 1.10 to 1.19 | 0.000 | 0.0000 | 0% | 0.79 | 1.10 to 1.20 |
| No industry ties declared | 10 | 1.18 | 1.07 to 1.29 | 1.09 to 1.27 | 0.082 | 0.0067 | 42% | 0.91 | 0.95 to 1.45 |
| No statement in PubMed record | 17 | 1.16 | 1.09 to 1.23 | 1.10 to 1.23 | 0.092 | 0.0085 | 64% | 0.98 | 0.95 to 1.43 |
| Risk difference, trials with counts (percentage points) | 36 | +6.5 | 5.1 to 7.9 | 5.1 to 7.9 | 2.5 points | 6.0 points² | 35% | 0.98 | 1.3 to 11.7 |
Hartung-Knapp intervals use the max(1, q) floor, so they are never narrower than the Wald interval. For the main analysis q = 0.976; without the floor the interval is 1.12 to 1.20. Prediction intervals use the Wald standard error, so the floor does not change them. I² is Higgins' I² from Cochran's Q. Software that reports a REML-based I² gives 46% for the main analysis. Unfloored intervals and both I² values for every row are in analysis/meta_adr_v2_results.json.
Caveats
- A stand-in outcome. ADR is an established quality measure: endoscopists with higher ADRs have patients with lower later risk of bowel cancer death. But these trials did not measure cancers or deaths, and the AGA rated the evidence for those outcomes as very low certainty [58].
- The extra finds are mostly small adenomas. More adenomas per colonoscopy were found with AI, driven mainly by small (5 mm or under) adenomas [4, 60], without a consistent gain in advanced lesions. Examples: advanced adenomas 21.3% vs 20.5% in a programme for people with a positive stool test [9]; advanced neoplasia 34.8% vs 34.6% [33]; and advanced ADR 18.5% vs 15.9%, not significant [45]. Results for sessile serrated lesions, another precancerous type, were mixed: higher with AI in one trial (22% vs 11%) [28], with no clear difference in others [16, 1, 29].
- There are costs as well as gains. More harmless polyps were removed in some trials (non-neoplastic resection 52.1% vs 35.0%) [32]. Some trials led to more intensive follow-up (more frequent surveillance recommended for 13.2% vs 7.3% of people with a positive stool test) [4]. Withdrawal time, the time spent inspecting the bowel, was longer with AI in several trials [1, 36].
- Possible deskilling. In four Polish centres, the ADR of colonoscopies done without AI fell from 28.4% (226 of 795) in the 3 months before AI was introduced to 22.4% (145 of 648) in the 3 months after [57]. That study was observational, so it does not show that AI caused the drop, and no pooled trial measured it.
- Hiding the arm is hard. An on-screen AI prompt is difficult to hide from the endoscopist. Only one trial used a sham system so that endoscopists and patients did not know the arm; its risk ratio was 1.23 (95% CI 1.02 to 1.49), close to the pooled result [49]. Because withdrawal time was often longer with AI, part of the gain may come from more careful inspection rather than the AI prompts themselves.
- Industry involvement is common. Many trials report device loans, consultancy or speaker fees from device makers. Others declare none, and 17 pooled trials had no statement in the PubMed record. Pooled results were similar across these groups. This is a standard caveat, not a judgment on any single trial.
- Possible small-study effects. Smaller and single-centre trials tended to show larger gains (see Results). Restricting to multicentre trials, or dropping the smallest trials, left the result much the same.
- Mostly abstract-level data. Most numbers come from PubMed abstracts. Open-access full texts were read for 10 included trials. Where only percentages and arm sizes were given, counts were back-calculated and checked against the reported percentage. Some denominators are randomized rather than analysed patients. One trial's abstract labels 1.94 a risk ratio, but that figure equals the odds ratio for its two reported rates (52.3% vs 36.1%). The risk ratio from those rates (1.45) was used instead.
- Possible overlap between trials. Two groups of pooled trials may share patients or centres. The first is three Italian multicentre trials with the same first author [20, 9, 1]; two of them (a stool-test screening trial and a surveillance trial) ran within an organised screening programme, so some people could appear in both. The second is three trials from the same hospital in Chengdu, China [51, 49, 47]. Their populations look distinct from the abstracts, but overlap cannot be ruled out without full texts.
- AI did the screening and coding. One AI screener and extractor did this work, with no human double-check. Every extracted number is matched by script to a verbatim quote in the fetched source.
- One database. Only PubMed was searched, so trials indexed elsewhere may be missed.
Methods
Design. A rapid non-systematic review, prepared with the Palo Alto Health Labs Co-Scientist. No protocol was registered and no formal risk-of-bias tool was applied.
Search. PubMed through NCBI E-utilities, searched 23 November 2026 and re-run the same evening with identical counts. Query 1 (107 records): ("colonoscopy"[tiab]) AND ("computer-aided detection"[tiab] OR "computer aided detection"[tiab] OR CADe[tiab] OR "artificial intelligence"[tiab] OR "deep learning"[tiab] OR "computer-aided"[tiab]) AND ("adenoma detection"[tiab] OR adenoma*[tiab]) AND (randomized controlled trial[pt] OR randomi*[ti]). Query 2 (213 records): (colonoscopy[tiab] OR colonoscopic[tiab]) AND ("computer-aided detection"[tiab] OR "computer aided detection"[tiab] OR CADe[tiab] OR "artificial intelligence"[tiab] OR "deep learning"[tiab] OR "computer-aided"[tiab] OR "computer-assisted"[tiab] OR "GI Genius"[tiab] OR ENDO-AID[tiab] OR CAD-EYE[tiab]) AND (polyp*[tiab] OR adenoma*[tiab] OR neoplasia[tiab]) AND (randomized controlled trial[pt] OR randomi*[tiab]) AND 2018:2026[dp]. After combining, there were 214 unique records (105 found by both).
Eligibility (screening/criteria.md). Primary reports of trials that randomized individual patients to parallel groups, compared real-time computer-aided polyp detection with the same colonoscopy without it, and reported ADR by arm. Add-ons were allowed only if both arms received them. Excluded:
- reviews and protocols (E1, 105);
- non-randomized studies (E2, 19);
- back-to-back (tandem) miss-rate designs (E3, 13). These measure how many adenomas a second look finds, not the share of patients with an adenoma, so they cannot give a per-patient ADR comparison;
- randomization by cluster, day or room (E4, 3);
- other AI tools, or arms differing by more than AI detection (E5, 17);
- no ADR outcome (E6, 4).
Included: 53 trials, 40 pooled and 13 summarized in the text. Version 2 moved two studies that version 1 had pooled to E2. Their full texts describe allocation by study period in one and historical controls in the other, not patient randomization. For every pooled trial, the randomization sentence was re-checked against the PubMed abstract (work_v2/design_check_v2.csv). Every decision is in screening/screening_log_v2.csv.
DOI check. DOIs were taken only from each PubMed record's own ArticleIdList (ELocationID as fallback) and checked against Crossref for title and first-author surname. An earlier parser had picked up reference-list DOIs for 71 of 214 records (16 of them included trials); all were corrected before analysis (doi_check_v2.csv). One included trial's DOI failed the first-author check and is cited by its PubMed link instead. One trial's PubMed title has a different subtitle from the publisher's; sources_v2.csv uses the publisher and Crossref title.
Analysis. The effect measure was the risk ratio of ADR, AI vs no AI, on the log scale. It came from 2×2 counts where available, otherwise from a reported risk ratio and its CI. The model was random-effects with REML between-trial variance (tau²). The main 95% CI is Hartung-Knapp with t on k−1 degrees of freedom and the max(1, q) floor. The unfloored interval, q, the Wald (z) 95% CI, tau, tau² and I² are reported alongside. I² is Higgins' I² from Cochran's Q. The 95% prediction interval is exp(μ ± t(0.975; k−2) × √(τ² + SE²)). Here μ is the pooled log risk ratio, τ² the REML between-trial variance, and SE the unfloored Wald standard error of μ, with no Hartung-Knapp scaling. The risk difference used the same model on the difference in ADR, for trials with patient counts in both arms. An exploratory random-effects meta-regression (R metafor, REML with Knapp-Hartung tests) related the log risk ratio to the control-arm ADR (per 10 percentage points). Its R² is metafor's pseudo-R², the proportional reduction in tau², truncated at zero. Small-study effects were examined with Egger's regression test, a rank correlation test and trim-and-fill (L0 estimator, descriptive only). They were also examined by re-pooling after removing the smallest trials (chosen before these analyses were run: the smallest trial, the 3 smallest, and all trials below the median total sample size), keeping multicentre trials only, and leaving out single-centre trials from China. Centre rule: a trial was multicentre if it enrolled patients at two or more separately named hospitals, and single-centre if it ran at one hospital, including two endoscopy units of one university hospital. Under this rule a trial run at two units of one university hospital is single-centre [28], and a trial run at two hospitals is multicentre even though one hospital contributed 98.1% of the cases [27]. Source of each coding: the abstract for 30 trials; the full text for 8 (7 where the abstract did not say, plus [28], checked for the rule); and, for 2 trials without open full text, the trial registry: one site for NCT04945044 [41] and five sites for NCT04260321 [46]. Per-trial sources: work_v2b/centre_sources_v2b.csv; final coding: work/adr_k40_inputs_v3.csv; subsets: analysis/design_subsets_v2b.py. Meta-regression and small-study analyses: work/sally_metareg_v3.md and work/sally_smallstudy_v3.md; the model with centre type was rechecked in analysis/m3_ranktest_v2b.R. Pooling required at least 3 comparable trials. A cross-check with DerSimonian-Laird in statsmodels agreed to 2 decimals. Code: analysis/meta_adr_v2.py; figures: figures_src/fig1_v2c.py, figures_src/figs23_v2c.py and figures_src/figs45_v2d.py. Meta-regression and small-study code: work/sally_metareg_v3.R and work/sally_smallstudy_v3.R (R 4.5.0, metafor 4.8.0). Software: Python 3.13.5, NumPy 2.5.5, SciPy 1.18.1, statsmodels 0.15.0, matplotlib 3.11.2.
References
Included trials (1–53):
- Spada 2026 (GENIAL-CO follow-up). Dig Liver Dis 2026. PMID 42838765. https://doi.org/10.1016/j.dld.2026.08.031
- Hüneburg 2026 (CADLY2). Lancet Gastroenterol Hepatol 2026. PMID 42462747. https://doi.org/10.1016/S2468-1253(26)00163-9
- Davila-Piñón 2026. Endoscopy 2026. PMID 42331035. https://doi.org/10.1055/a-2895-1613
- Hsu 2026. JAMA Netw Open 2026. PMID 41984482. https://doi.org/10.1001/jamanetworkopen.2026.4881
- Santos 2026. Clin Endosc 2026. PMID 41943654. https://doi.org/10.5946/ce.2025.276
- Gimeno-García 2026. Gastroenterol Hepatol 2026. PMID 41935575. https://doi.org/10.1016/j.gastrohep.2026.502757 (not pooled)
- Lux 2026. NPJ Digit Med 2026. PMID 41922731. https://doi.org/10.1038/s41746-026-02576-8
- Robles de la Osa 2026. Colorectal Dis 2026. PMID 41854124. https://doi.org/10.1111/codi.70426 (not pooled)
- Spada 2026 (FIT programme). United European Gastroenterol J 2026. PMID 41563802. https://doi.org/10.1002/ueg2.70176
- Leggett 2026. Gastrointest Endosc 2026. PMID 41548723. https://doi.org/10.1016/j.gie.2025.12.290 (not pooled)
- Dos Santos 2025. Rev Gastroenterol Peru 2025. PMID 41530068. https://pubmed.ncbi.nlm.nih.gov/41530068/ (not pooled)
- Kader 2025 (EAGLE). NPJ Digit Med 2025. PMID 41449203. https://doi.org/10.1038/s41746-025-02270-1
- Jeon 2026. Gut Liver 2026. PMID 41306099. https://doi.org/10.5009/gnl250369 (PubMed: https://pubmed.ncbi.nlm.nih.gov/41306099/)
- Schauer 2026. J Gastroenterol Hepatol 2026. PMID 41287376. https://doi.org/10.1111/jgh.70145
- Xu 2025. Surg Endosc 2025. PMID 40897874. https://doi.org/10.1007/s00464-025-12080-x
- Cheng 2026. Endoscopy 2026. PMID 40789324. https://doi.org/10.1055/a-2681-5544
- Yabuuchi 2025. Dig Endosc 2025. PMID 40611562. https://doi.org/10.1111/den.15086
- Zimmermann-Fraedrich 2026. Am J Gastroenterol 2026. PMID 40293139. https://doi.org/10.14309/ajg.0000000000003500 (not pooled)
- Alali 2025. J Clin Med 2025. PMID 39860586. https://doi.org/10.3390/jcm14020581
- Spada 2025 (ACCENDO-Colo). Dig Liver Dis 2025. PMID 39814659. https://doi.org/10.1016/j.dld.2024.12.023
- Park 2024. Sci Rep 2024. PMID 39455850. https://doi.org/10.1038/s41598-024-77079-1
- Seager 2024 (COLO-DETECT). Lancet Gastroenterol Hepatol 2024. PMID 39153491. https://doi.org/10.1016/S2468-1253(24)00161-4
- Ortiz 2024 (TIMELY). Lancet Gastroenterol Hepatol 2024. PMID 39033774. https://doi.org/10.1016/S2468-1253(24)00187-0 (not pooled)
- Maas 2024 (Endoscopy). Endoscopy 2024. PMID 38749482. https://doi.org/10.1055/a-2328-2844
- Thiruvengadam 2024. Clin Gastroenterol Hepatol 2024. PMID 38437999. https://doi.org/10.1016/j.cgh.2024.02.021 (not pooled)
- Maas 2024 (Lancet Digit Health). Lancet Digit Health 2024. PMID 38395537. https://doi.org/10.1016/S2589-7500(23)00242-X (not pooled)
- Lui 2024 (AI vs HD arms). Am J Gastroenterol 2024. PMID 38305278. https://doi.org/10.14309/ajg.0000000000002684
- Schöler 2024. BMJ Open Gastroenterol 2024. PMID 38290758. https://doi.org/10.1136/bmjgast-2023-001247
- Desai 2024. Am J Gastroenterol 2024. PMID 38235741. https://doi.org/10.14309/ajg.0000000000002664
- Miyaguchi 2024. Endoscopy 2024. PMID 38191000. https://doi.org/10.1055/a-2239-8145
- Tiankanon 2024. Endoscopy 2024. PMID 37963587. https://doi.org/10.1055/a-2210-7999 (not pooled)
- Lau 2024. Clin Gastroenterol Hepatol 2024. PMID 37918685. https://doi.org/10.1016/j.cgh.2023.10.019
- Mangas-Sanjuan 2023. Ann Intern Med 2023. PMID 37639723. https://doi.org/10.7326/M22-2619
- Karsenti 2023 (COLO-GENIUS). Lancet Gastroenterol Hepatol 2023. PMID 37269872. https://doi.org/10.1016/S2468-1253(23)00104-8
- Engelke 2023. Scand J Gastroenterol 2023. PMID 37191195. https://doi.org/10.1080/00365521.2023.2212309 (not pooled)
- Wei 2023 (AI-SEE). Am J Gastroenterol 2023. PMID 36892545. https://doi.org/10.14309/ajg.0000000000002239
- Lachter 2023. IGIE 2023. PMID 41647953. https://doi.org/10.1016/j.igie.2023.01.013
- Nakashima 2023. Digestion 2023. PMID 36599306. https://doi.org/10.1159/000528085
- Hüneburg 2023. United European Gastroenterol J 2023. PMID 36571259. https://doi.org/10.1002/ueg2.12354
- Ahmad 2023 (AI-DETECT). Endoscopy 2023. PMID 36509103. https://doi.org/10.1055/a-1966-0661
- Gimeno-García 2023. Gastrointest Endosc 2023. PMID 36228695. https://doi.org/10.1016/j.gie.2022.09.029
- Aniwan 2023. Gastrointest Endosc 2023. PMID 36220382. https://doi.org/10.1016/j.gie.2022.09.023 (not pooled)
- Xu 2023. Clin Gastroenterol Hepatol 2023. PMID 35863686. https://doi.org/10.1016/j.cgh.2022.07.006
- Shaukat 2022 (SKOUT). Gastroenterology 2022. PMID 35643173. https://doi.org/10.1053/j.gastro.2022.05.028
- Rondonotti 2022. Endoscopy 2022. PMID 35545122. https://doi.org/10.1055/a-1849-6878
- Repici 2022 (non-expert trial). Gut 2022. PMID 34187845. https://doi.org/10.1136/gutjnl-2021-324471
- Liu 2020 (Ther Adv Gastroenterol). Ther Adv Gastroenterol 2020. PMID 33403003. https://doi.org/10.1177/1756284820979165
- Repici 2020. Gastroenterology 2020. PMID 32371116. https://doi.org/10.1053/j.gastro.2020.04.062
- Wang 2020 (CADe-DB, sham-controlled). Lancet Gastroenterol Hepatol 2020. PMID 31981517. https://doi.org/10.1016/S2468-1253(19)30411-X
- Liu 2020 (Saudi J Gastroenterol). Saudi J Gastroenterol 2020. PMID 31898644. https://pubmed.ncbi.nlm.nih.gov/31898644/ (not pooled)
- Wang 2019. Gut 2019. PMID 30814121. https://doi.org/10.1136/gutjnl-2018-317500
- Yao 2022 (CADe vs control arms). Endoscopy 2022. PMID 34823258. https://doi.org/10.1055/a-1706-6174
- Vilkoite 2023. Diagnostics (Basel) 2023. PMID 36832189. https://doi.org/10.3390/diagnostics13040701 (not pooled)
Other studies cited:
- Codesido-Prado 2026 (day-randomized surveillance trial). Endoscopy 2026. PMID 42365851. https://doi.org/10.1055/a-2903-9497
- Dominitz 2026 (Veterans Affairs facility-randomized study). Gastroenterology 2026. PMID 42250891. https://doi.org/10.1053/j.gastro.2026.05.018
- Ladabaum 2023 (pragmatic implementation study). Gastroenterology 2023. PMID 36528131. https://doi.org/10.1053/j.gastro.2022.12.004
- Budzyń 2025 (observational deskilling study). Lancet Gastroenterol Hepatol 2025. PMID 40816301. https://doi.org/10.1016/S2468-1253(25)00133-5
- Sultan 2025 (AGA living clinical practice guideline). Gastroenterology 2025. PMID 40121061. https://doi.org/10.1053/j.gastro.2025.01.002
- Soleymanjahi 2024 (meta-analysis). Ann Intern Med 2024. PMID 39531400. https://doi.org/10.7326/ANNALS-24-00981
- Makar 2025 (meta-analysis). Gastrointest Endosc 2025. PMID 39216648. https://doi.org/10.1016/j.gie.2024.08.033
Corrections
None yet.
Found an error? Email hello@paloaltohealthlabs.com.
More from the Palo Alto Health Labs Co-Scientist · All posts
Have your own questions? Sign up for the Beta
Tell us the health AI or digital health question you would want the Co-Scientist to review, and join the beta list.
Scope it in 30 minutes
Book a scoping call and we'll name the smallest engagement that moves you forward. Most teams start with the fixed-scope Initial Audit.
Book a 30-min scoping callPrefer to write? Send a short note.